background image

MO290-en-GB_V3.6   4/15 

Condensation Mode 

The Condensation feature alerts the used when the surface temperature as measured by the IR 
thermometer is close to or has reached the Dew Point temperature. 

1. 

Press the power 

 button to turn the meter on. 

2. 

Simultaneously press the MOIST/REL and RH buttons. The “COND” icon will appear. 

3. 

Point the meter at a surface, press the IRT button to measure the surface temperature. The 
small display will indicate the IR surface temperature and the large display will indicate the 
difference between the IR temperature and the Dew Point temperature. 

4. 

The meter will then report the potential for condensation on that surface in the following 
manner 

  If the temperature of the IRT is more than 14°C (25°F) above the Dew Point, the 

temperature difference shall be displayed, with no other warning. 

  If the temperature of the IRT is 3-14°C (5-25°F) above the Dew Point, the temperature 

difference shall be displayed, along with a standard Condensation Indicator icon.  The meter 
shall beep once to confirm that the reading is in the risk area. 

  If the temperature of the IRT is less than 3°C (5°F) above the Dew Point, the temperature 

difference shall be displayed, along with a flashing Condensation Indicator icon.  The meter 
shall beep twice to confirm that the reading is in the high-risk area. 

5. 

Press the RH button to exit the mode. 

Vapor Pressure Mode 

1.  With the Condensation mode active, press the MODE button to display the Vapor Pressure in 

mBAR (°F) or kPa (°C). 

2.  Press the MODE button to exit the Vapor Pressure mode. 

Alarm High and Low Limit Setting 

High and Low alarm points can be set for Humidity and Moisture measurements. 
Humidity Alarm Set Procedure: 
1. 

With RH% displayed, simultaneously press the RH and MODE buttons. 

2. 

The “HIGH” icon will appear on the display. 

3. Press 

the 

or 

button to set the high limit desired. 

4. 

Press the ALARM SET button to save the value and proceed to the LOW set value. 

5. 

With the “LOW” icon in the display, Press the 

or 

button to set the low limit desired. 

6. 

Press the ALARM SET button to save the value and to return to the normal mode. 

7. 

If the humidity measurement is lower than the low alarm setting or higher than the high alarm 
setting, the meter will beep once every second. 

Moisture Alarm Set Procedure: 

1. 

With MOIST displayed, simultaneously press the MOIST/REL and MODE buttons. 

2. 

The “HIGH” icon will appear on the display. 

3. Press 

the 

or 

button to set the high limit desired. 

4. 

Press the ALARM SET button to save the value and proceed to the LOW set value. 

5. 

With the “LOW” icon in the display, Press the 

or 

button to set the low limit desired. 

6. 

Press the ALARM SET button to save the value and to return to the normal mode. 

7. 

If the moisture measurement is higher than the LOW alarm setting, the meter will beep once 
every second 

8. 

If the moisture measurement is higher than the HIGH alarm setting, the meter will beep 
continuously. 

Change Temperature units from F to C or C to F 

1. 

Press the power 

 button to turn the meter on. 

2. 

Press the IRT button to turn on the IR thermometer and then release the button. 

3. Press 

the 

or 

button to set the desired temperature unit 

Отзывы: