Extech Instruments HD600 Скачать руководство пользователя страница 8

HD600-EU-EN v1.3   07/13 

USB PC Interface 

The meter has a built-in USB port for use with the supplied data acquisition software. The software 
allows the user to download stored data, view, save, export, and print readings from the sound level 
meter.  

SOFTWARE INSTALLATION 

Refer to the documentation included with the software for complete details on installation and 
operation of the application program and the USB driver software. 

SOFTWARE COMMUNICATION 

1. 

Connect the meter to the PC using the supplied USB cable. 

2. 

Turn the meter ON and press the SETUP button. 

3. 

Launch the application program. 

4. 

Select the COM port which has the CP210X driver installed. 

5. 

Data will appear on the pc screen when communication is established. 

Specifications 

Applicable Standards

 

IEC61672-1: 2002 Class 2; IEC60651: 1979 Type 2; ANSI S1.4:1983 

Type 2, 

 

Accuracy 

±1.4dB (under reference conditions)

 

Frequency Range 

31.5Hz to 8kHz

 

Dynamic Range 

50dB

 

Frequency Weighting 

A and C

 

Time response 

Fast (125ms) and Slow (1 second)

 

Measurement Ranges 

30 to 80dB, 50 to 100dB, 80 to 130dB and autoranging (30 to 130dB) 

Memory 

20,000 records with date and time 

Microphone 

½” electret condenser 

Calibration 

Requires external calibrator 

Display  

4 digit LCD with bargraph and backlighting

 

Display update rate 

2 times/second 

Range indicators 

OVER” and “UNDER” range indication 

Battery life 

30 hours (approximately)

 

Power supply 

One 9V battery (NEDA1604 or equivalent) or 12V/1A AC adaptor

 

Auto Power Off 

After approx. 15 minutes of inactivity with disable 

Analog Outputs

 

AC: 1Vrms full scale; Output impedance: 100

 

DC: 10mV/1dB; Output impedance: 1k

 

Operating conditions

 

32 to 104

o

F (0 to 40

o

C); 10% to 90% relative humidity 

Storage conditions

 

14 to 140

o

F (-10 to 60

o

C); 10% to75% relative humidity 

Dimensions

 

10.9 x 3 x 1.97” (278 x 76 x 50mm) 

Weight 

12.35 oz. (350g) 

 

Содержание HD600

Страница 1: ...User s Guide Digital Datalogging Sound Level Meter Model HD600 ...

Страница 2: ...nd software enable the user to download and analyze measured data This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Safety Read the following safety information carefully before attempting to operate or service the meter Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired Environmen...

Страница 3: ...d Mount rear Battery Compartment rear SIDE PANEL 13 External 9VDC power adaptor 14 USB pc port 15 DC AC analog output 16 CAL potentiometer DISPLAY Symbol Function MAX Maximum hold MIN Minimum hold OVER Over range UNDER Under range FAST Fast response SLOW Slow response dBA A Weighting dBC C Weighting REC Recording data AUTO Auto range selection FULL Memory full HOLD Data hold Auto Power Off enabled...

Страница 4: ...n to enter the minutes adjustment screen Press the LEVEL button to adjust the minutes to the current time 4 Press the SETUP button to enter the hour adjustment screen Press the LEVEL button to adjust the hour to the current time h P indicates PM and h A indicates AM 5 Press the SETUP button to enter the date adjustment screen Press the LEVEL button to adjust the date to the current day 6 Press the...

Страница 5: ...nd bargraph displays indicate the sound level measurement 6 To turn the meter off press and hold the button for 3 seconds LEVEL The LEVEL button is used to select the measurement range Press this button to step through the ranges as indicated on the display 30 to 80 50 to 100 80 to 130 or 30 to 130 autoranging A C FREQUENCY WEIGHTING Press the A C button to select A or C frequency weighting With A...

Страница 6: ...hold the button while turning the meter ON 0001 and Int will appear in the display 2 Press the LEVEL button to set the sample interval from once per second to once per 59 seconds 3 Press the HOLD button to save the interval setting and exit the setting mode Recording readings 1 Press the REC button to begin recording The REC icon will appear on the display 2 Press the REC button again to stop reco...

Страница 7: ...irectives 1 Turn the meter ON 2 Put the meter in the A weighting mode 3 Put the meter in the SLOW response mode 4 Place the calibrator onto the microphone 5 Turn the calibrator ON 6 Adjust the meter s CAL potentiometer located on the side panel so that the meter s display matches the output of the calibrator typically 94dB or 114dB Battery Replacement 1 The battery icon will appear on the display ...

Страница 8: ...Type 2 ANSI S1 4 1983 Type 2 Accuracy 1 4dB under reference conditions Frequency Range 31 5Hz to 8kHz Dynamic Range 50dB Frequency Weighting A and C Time response Fast 125ms and Slow 1 second Measurement Ranges 30 to 80dB 50 to 100dB 80 to 130dB and autoranging 30 to 130dB Memory 20 000 records with date and time Microphone electret condenser Calibration Requires external calibrator Display 4 digi...

Страница 9: ... Pneumatic drill 50 Freight train 100 Vacuum cleaner 10 Speech 1 Large store Large Office Small office Residence Night residential area Whisper 5 0 Sound studio Threshold of hearing Copyright 2013 FLIR Systems Inc All rights reserved including the right of reproduction in whole or in part in any form www extech com Please visit our website for more information www alpine components co uk ...

Отзывы: