MEASUREMENT CONSIDERATIONS 

1. 

Wind blowing across the microphone increases the noise measurement. Use the supplied 
windscreen to cover the microphone when applicable. 

2. 

Calibrate the instrument before each use if possible. Especially if the meter has not been used 
for a long period of time. 

3. 

Do not store or operate the instrument in areas of high temperature or humidity. 

4. 

Keep meter and microphone dry. 

5. 

Avoid severe vibration. Protect the meter from impact. Do not drop it. Transport the meter in 
the supplied case. 

6. 

Remove the battery when the meter is to be stored for long periods of time. 

BASIC OPERATION 

1. 

Turn the meter ON by pressing the 

  button. 

2. 

Select ‘A’ or ‘C” frequency weighting by pressing the A/C button. 

3. 

Select Fast or Slow response time by pressing the FAST/SLOW button. 

4. 

Use the LEVEL button to select the appropriate range. The range is shown near the top of the 
display. Use a range that puts the sound level reading in the center of the range. If the OVER 
or UNDER icons appear on the display, select a new range if possible. 

5. 

The numeric and bargraph displays indicate the sound level measurement. 

6. 

To turn the meter off, press and hold the 

  button for 3 seconds. 

LEVEL 

The LEVEL button is used to select the measurement range. Press this button to step through the 
ranges as indicated on the display (30 to 80, 50 to 100, 80 to 130, or 30 to 130 autoranging) 

A/C FREQUENCY WEIGHTING 

Press the A/C button to select ‘A’ or ‘C’ frequency weighting. With ‘A’ weighting selected, the 
frequency response of the meter is similar to the response of the human ear. ‘A’ weighting is 
commonly used for environmental or hearing conservation programs such as OSHA regulatory 
testing and noise ordinance law enforcement. ‘C’ weighting is a much flatter response and is 
suitable for the sound level analysis of machines, engines, etc. Most noise measurements are 
performed using 'A' Weighting and SLOW Response. 

FAST/SLOW RESPONSE TIME 

Use the FAST/SLOW button to select a FAST (125 ms) or a SLOW (1 second) response time. 
Select FAST to capture noise peaks and noises that occur very quickly. Select the SLOW response 
to monitor a sound source that has a consistent noise level or to average quickly changing levels. 
Select SLOW response for most applications. 

HOLD 

Press the HOLD button “freeze” the current reading in the display. Press the button again to resume 
normal operation. 

 

 

Содержание HD600

Страница 1: ...User s Guide Digital Datalogging Sound Level Meter Model HD600 ...

Страница 2: ...nd software enable the user to download and analyze measured data This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Safety Read the following safety information carefully before attempting to operate or service the meter Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired Environmen...

Страница 3: ...d Mount rear Battery Compartment rear SIDE PANEL 13 External 9VDC power adaptor 14 USB pc port 15 DC AC analog output 16 CAL potentiometer DISPLAY Symbol Function MAX Maximum hold MIN Minimum hold OVER Over range UNDER Under range FAST Fast response SLOW Slow response dBA A Weighting dBC C Weighting REC Recording data AUTO Auto range selection FULL Memory full HOLD Data hold Auto Power Off enabled...

Страница 4: ...n to enter the minutes adjustment screen Press the LEVEL button to adjust the minutes to the current time 4 Press the SETUP button to enter the hour adjustment screen Press the LEVEL button to adjust the hour to the current time h P indicates PM and h A indicates AM 5 Press the SETUP button to enter the date adjustment screen Press the LEVEL button to adjust the date to the current day 6 Press the...

Страница 5: ...nd bargraph displays indicate the sound level measurement 6 To turn the meter off press and hold the button for 3 seconds LEVEL The LEVEL button is used to select the measurement range Press this button to step through the ranges as indicated on the display 30 to 80 50 to 100 80 to 130 or 30 to 130 autoranging A C FREQUENCY WEIGHTING Press the A C button to select A or C frequency weighting With A...

Страница 6: ...hold the button while turning the meter ON 0001 and Int will appear in the display 2 Press the LEVEL button to set the sample interval from once per second to once per 59 seconds 3 Press the HOLD button to save the interval setting and exit the setting mode Recording readings 1 Press the REC button to begin recording The REC icon will appear on the display 2 Press the REC button again to stop reco...

Страница 7: ...irectives 1 Turn the meter ON 2 Put the meter in the A weighting mode 3 Put the meter in the SLOW response mode 4 Place the calibrator onto the microphone 5 Turn the calibrator ON 6 Adjust the meter s CAL potentiometer located on the side panel so that the meter s display matches the output of the calibrator typically 94dB or 114dB Battery Replacement 1 The battery icon will appear on the display ...

Страница 8: ...Type 2 ANSI S1 4 1983 Type 2 Accuracy 1 4dB under reference conditions Frequency Range 31 5Hz to 8kHz Dynamic Range 50dB Frequency Weighting A and C Time response Fast 125ms and Slow 1 second Measurement Ranges 30 to 80dB 50 to 100dB 80 to 130dB and autoranging 30 to 130dB Memory 20 000 records with date and time Microphone electret condenser Calibration Requires external calibrator Display 4 digi...

Страница 9: ... Pneumatic drill 50 Freight train 100 Vacuum cleaner 10 Speech 1 Large store Large Office Small office Residence Night residential area Whisper 5 0 Sound studio Threshold of hearing Copyright 2013 FLIR Systems Inc All rights reserved including the right of reproduction in whole or in part in any form www extech com Please visit our website for more information www alpine components co uk ...

Отзывы: