HD600-EU-EN v1.3   07/13 

MAX / MIN  

In the MAX/MIN mode the meter will display and hold either the maximum or minimum reading. The 
display will update only when the measured value exceeds the value presently in the display. 

1. 

Press the MAX/MIN button and the MAX icon will appear on the display. The reading displayed 
is the highest reading encountered since the MAX mode was entered.  

2. 

Press the MAX/MIN button again. The MIN icon will appear on the display. The reading 
displayed is the lowest reading encountered since the MIN mode was entered. 

3. 

Press the MAX/MIN button again to exit the MAX / MIN display mode.  

RECORDING DATA 

The meter can store up to 20,000 readings at an interval rate of 1 to 59 seconds. Each recording 
session is saved as a data set and each record is saved with a date and time stamp. This data can 
be downloaded using the supplied software. 

Setting the interval time 

1. 

Press and hold the       button while turning the meter ON. 

0001

 and 

Int

 will appear in the 

display. 

2. 

Press the LEVEL button to set the sample interval from once per second to once per 59 
seconds. 

3. 

Press the HOLD button to save the interval setting and exit the setting mode. 

Recording readings 

1. 

Press the REC button to begin recording. The 

REC

 icon will appear on the display. 

2. 

Press the REC button again to stop recording. 

3. 

Auto Power OFF is disabled when the record function is active. 

Clearing Stored readings 

1. 

Turn the meter OFF. 

2. 

Press and hold the REC button while turning the meter ON. 

3. 

When         (clear) appears on the display, release the REC button. 

4. 

All of the readings that were stored in memory are now erased.  

BACKLIGHT 

Press the     button to turn the LCD backlight on or off. To conserve battery life, the backlight will 
automatically turn off after approximately 30 seconds. 

AUTO POWER OFF 

The meter will automatically shut off after approximately 15 minutes of inactivity. Press the SETUP 

button to disable the auto-power-off feature. The 

 icon in the display indicates that the auto power 

off feature is active. 

 

 

Содержание HD600

Страница 1: ...User s Guide Digital Datalogging Sound Level Meter Model HD600 ...

Страница 2: ...nd software enable the user to download and analyze measured data This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Safety Read the following safety information carefully before attempting to operate or service the meter Use the meter only as specified in this manual otherwise the protection provided by the meter may be impaired Environmen...

Страница 3: ...d Mount rear Battery Compartment rear SIDE PANEL 13 External 9VDC power adaptor 14 USB pc port 15 DC AC analog output 16 CAL potentiometer DISPLAY Symbol Function MAX Maximum hold MIN Minimum hold OVER Over range UNDER Under range FAST Fast response SLOW Slow response dBA A Weighting dBC C Weighting REC Recording data AUTO Auto range selection FULL Memory full HOLD Data hold Auto Power Off enabled...

Страница 4: ...n to enter the minutes adjustment screen Press the LEVEL button to adjust the minutes to the current time 4 Press the SETUP button to enter the hour adjustment screen Press the LEVEL button to adjust the hour to the current time h P indicates PM and h A indicates AM 5 Press the SETUP button to enter the date adjustment screen Press the LEVEL button to adjust the date to the current day 6 Press the...

Страница 5: ...nd bargraph displays indicate the sound level measurement 6 To turn the meter off press and hold the button for 3 seconds LEVEL The LEVEL button is used to select the measurement range Press this button to step through the ranges as indicated on the display 30 to 80 50 to 100 80 to 130 or 30 to 130 autoranging A C FREQUENCY WEIGHTING Press the A C button to select A or C frequency weighting With A...

Страница 6: ...hold the button while turning the meter ON 0001 and Int will appear in the display 2 Press the LEVEL button to set the sample interval from once per second to once per 59 seconds 3 Press the HOLD button to save the interval setting and exit the setting mode Recording readings 1 Press the REC button to begin recording The REC icon will appear on the display 2 Press the REC button again to stop reco...

Страница 7: ...irectives 1 Turn the meter ON 2 Put the meter in the A weighting mode 3 Put the meter in the SLOW response mode 4 Place the calibrator onto the microphone 5 Turn the calibrator ON 6 Adjust the meter s CAL potentiometer located on the side panel so that the meter s display matches the output of the calibrator typically 94dB or 114dB Battery Replacement 1 The battery icon will appear on the display ...

Страница 8: ...Type 2 ANSI S1 4 1983 Type 2 Accuracy 1 4dB under reference conditions Frequency Range 31 5Hz to 8kHz Dynamic Range 50dB Frequency Weighting A and C Time response Fast 125ms and Slow 1 second Measurement Ranges 30 to 80dB 50 to 100dB 80 to 130dB and autoranging 30 to 130dB Memory 20 000 records with date and time Microphone electret condenser Calibration Requires external calibrator Display 4 digi...

Страница 9: ... Pneumatic drill 50 Freight train 100 Vacuum cleaner 10 Speech 1 Large store Large Office Small office Residence Night residential area Whisper 5 0 Sound studio Threshold of hearing Copyright 2013 FLIR Systems Inc All rights reserved including the right of reproduction in whole or in part in any form www extech com Please visit our website for more information www alpine components co uk ...

Отзывы: