EX622-en-GB_v2.8 4/15
13
Specifications
Function Range
Resolution
Accuracy
(% of r digits)
AC Current
50/60 Hz
True RMS
400.0 AAC
0.1A
±(1.5% + 5 digits)
40.00 AAC
0.01A
AC/DC
A
Current
400.00
A 0.01
A
DC: ±(1.0% + 2 digits)
AC: ±(1.5% + 2 digits)
4000.0
A 0.1
A
AC Voltage
50/60 Hz
True RMS
400.0 mVAC
0.1mV
±(1.0% + 20 digits)
4.000 VAC
0.001V
±(2.0% + 5 digits)
40.00 VAC
0.01V
400.0 VAC
0.1V
600 VAC
1V
DC Voltage
400.00 mVDC
0.01mV
±(0.1% + 6 digits)
4.0000 VDC
0.0001V
40.000 VDC
0.001V
400.00 VDC
0.01V
600.0 VDC
0.1V
±(1.0% + 2 digits)
Resistance
400.00
0.01
±(0.8% + 20 digits)
4.0000k
0.0001k
±(0.8% + 4 digits)
40.000k
0.001k
400.00k
0.01k
4.0000M
0.0001M
40.000M
0.001M
±(2.5% + 10 digits)
Capacitance
400.00nF
0.01nF
±(5.0% + 40 digits)
4000.0nF
0.1nF
±(3.0% + 10 digits)
400.00µF
0.01µF
±(3.5% + 10 digits)
4000.0µF 0.1µF
±(5.0% + 10 digits)
40.000mF 0.001mF
Frequency
(clamp)
400.00Hz
0.01Hz
±(1.0% + 3 digits)
Sensitivity: 5Arms minimum