background image

EX622-en-GB_v2.8   4/15 

11

Data Hold 

To freeze the LCD reading, press the 

HOLD

 button. While data hold is active, the 

HOLD

 icon 

appears on the LCD. Press the

 HOLD

 button again to return to normal operation. 

MAX/MIN  

1. Press 

the 

MAX/MIN

 button to activate the MAX/MIN recording mode. The display icon "

MAX

will appear. The meter will begins recording and displaying the maximum value measured. 

2. Press 

the 

MAX/MIN

 button and “

MIN

” will appear. The meter will display the minimum value 

measured during the recording session. 

3. Press 

the 

MAX/MIN

 button and “

MAX

 

MIN

” will appear. The meter will display the present 

reading, but will continue to update and store the max and min readings. 

4.  To exit MAX/MIN mode press and hold the 

MAX/MIN

 button for 2 seconds. 

 

Peak Hold 

When ACA or ACV is selected, pressing the 

PEAK

 button enables the peak capture circuit. The 

meter will now capture and display the maximum and minimum peaks of the waveform. 

RANGE 

In the Voltage, Resistance, Capacitance, Frequency or uA function the meter automatically selects 
the best range for the measurements being made. For measurement situations requiring that a 
range be manually selected, perform the following: 

1. Press 

the

 RANGE

 button. The “

AUTO”

 display icon will turn off. 

2. Press 

the 

RANGE

 key to step through the available ranges. Observe the decimal point and 

units displayed until the preferred range is located. 

3. 

To exit the Manual Ranging mode and return to Autoranging, press and hold the

 RANGE

 key 

for 2 seconds. 

LCD Backlight 

The LCD is equipped with backlighting for easier viewing, especially in dimly lit areas. Press and 
hold the 

HOLD/

 button for 2 seconds to turn the backlight on. The backlight will automatically turn 

off after 30 seconds.  

Automatic Power OFF with Disable 

In order to conserve battery life, the meter will automatically turn off after approximately 30 minutes. 
To turn the meter on again, turn the function switch to the OFF position and then to the desired 
function position. 
To disable APO: 

1.  From the OFF position, hold the MODE button and rotate the FUNCTION switch to a 

measurement function. 

2. 

 will appear in the display 

3.  Release the MODE button 
4.  APO is now disabled (APO icon is off) and will be reset when the Function switch is returned to 

the OFF position. 

Low battery indication 

When the 

 icon appears in the display, the battery should be replaced. Refer to the battery 

replacement procedure in the maintenance section. 

 

   www.ponpe.com

contact:

[email protected]

Содержание EX622

Страница 1: ...User s Guide 400Amp True RMS AC Clamp Meter with IR Thermometer Model EX622 www ponpe com contact sale ponpe com ...

Страница 2: ...injury This CAUTION symbol indicates a potentially hazardous situation which if not avoided may result damage to the product PER IEC1010 OVERVOLTAGE INSTALLATION CATEGORY OVERVOLTAGE CATEGORY I Equipment of OVERVOLTAGE CATEGORY I is equipment for connection to circuits in which measures are taken to limit the transient overvoltages to an appropriate low level Note Examples include protected electr...

Страница 3: ...trical outlets can be difficult and misleading because of the uncertainty of connection to the recessed electrical contacts Other means should be used to ensure that the terminals are not live If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired This device is not a toy and must not reach children s hands It contains hazard...

Страница 4: ...Voltage Detector 17 NCV LED indicator 18 Input shutter Display icons Description HOLD Data Hold APO Auto Power Off AUTO Autoranging P Peak Hold DC Direct Current AC Alternating Current MAX Max reading MIN Min reading Low battery ZERO DCA or CAP zero mV or V Milli volts or Volts Voltage Ohms Resistance A Amperes Current F Farad Capacitance Hz Hertz Frequency Duty Ratio o F and o C Fahrenheit and Ce...

Страница 5: ...re measurements Non Contact Voltage Detector WARNING Risk of Electrocution Before use always test the Voltage Detector on a known live circuit to verify proper operation 1 Rotate the Function switch to any measurement position 2 Place the detector probe tip on the conductor to be tested 3 If AC voltage is present the NCV detector light will turn on with a steady red light NOTE The conductors in el...

Страница 6: ...t is being switched ON or OFF Large voltage surges may occur that can damage the meter 1 Slide the input shutter to the up position 2 Rotate the function switch to the V position 3 Press the MODE button to select AC or DC Voltage 4 Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive V jack 5 Touch the black test probe tip to the ...

Страница 7: ...urements 2 Press the MODE button to select continuity 3 Touch the test probe tips across the circuit or component under test 4 If the resistance is 50 a tone will sound Diode Test 1 Connect as described for resistance measurements 2 Press the MODE button to select diode test 3 Touch the test probe tips to the diode or semiconductor junction under test Note the meter reading 4 Reverse the test lead...

Страница 8: ...ature removes stray test lead capacitance to improve the accuracy of low value capacitance measurements To perform a zero Press and hold the MODE ZERO button for two beeps The display will zero The offset value is now stored and is removed from all measurements Frequency and Duty Ratio Measurements 1 Slide the input shutter to the up position 2 Rotate the function switch to the Hz Position 3 Inser...

Страница 9: ... the red test probe tip to the positive side of the break 7 Turn circuit power on 8 Read the current value in the display Type K Temperature Measurements 1 Slide the input shutter to the down position 2 Rotate the function switch to the TYPE K temperature position 3 Press the MODE button to select F or C 4 Insert the Temperature Probe s into the T1 and or T2 type k sockets 5 Place the temperature ...

Страница 10: ...irect the laser pointer at an eye Low power visible lasers do not normally present a hazard but may present some potential for hazard if viewed directly for extended periods of time IR Spot to Distance Diagram The 8 1 spot to distance ratio determines the size of the measured surface area with respect to the distance the meter is held away from the surface IR Measurement Notes 1 The object under t...

Страница 11: ...e be manually selected perform the following 1 Press the RANGE button The AUTO display icon will turn off 2 Press the RANGE key to step through the available ranges Observe the decimal point and units displayed until the preferred range is located 3 To exit the Manual Ranging mode and return to Autoranging press and hold the RANGE key for 2 seconds LCD Backlight The LCD is equipped with backlighti...

Страница 12: ...of used batteries or rechargeable batteries in household waste As consumers users are legally required to take used batteries to appropriate collection sites the retail store where the batteries were purchased or wherever batteries are sold Disposal Do not dispose of this instrument in household waste The user is obligated to take end of life devices to a designated collection point for the dispos...

Страница 13: ...01V 400 0 VAC 0 1V 600 VAC 1V DC Voltage 400 00 mVDC 0 01mV 0 1 6 digits 4 0000 VDC 0 0001V 40 000 VDC 0 001V 400 00 VDC 0 01V 600 0 VDC 0 1V 1 0 2 digits Resistance 400 00 0 01 0 8 20 digits 4 0000k 0 0001k 0 8 4 digits 40 000k 0 001k 400 00k 0 01k 4 0000M 0 0001M 40 000M 0 001M 2 5 10 digits Capacitance 400 00nF 0 01nF 5 0 40 digits 4000 0nF 0 1nF 3 0 10 digits 400 00µF 0 01µF 3 5 10 digits 4000...

Страница 14: ... 2 digits Pulse width 100 s to 100ms Frequency 5Hz to 150kHz Temperature Type K 58 to 4 F 50 to 19 C 0 1 1000 1 1000 7 C 13 F 4 to 31 F 20 to 1 C 1 0 1 C 2 F 32 F 0 C 1 C 2 F 33 to 211 F 1 to 100 C 1 0 1 C 2 F 212 to 718 F 101 to 399 C 1 5 2 C 3 F 719 to 1832 F 400 to 1000 C 2 5 4 C 7 F Specification does not include probe accuracy Temperature IR 58 to 4 F 50 to 20 C 0 1 5 C 9 F 4 to 31 F 20 to 1 ...

Страница 15: ...onse True rms AAC and VAC Crest Factor 3 0 in 40A and 400A ranges 1 4 in 1000A range 50 60Hz and 5 to 100 of range Operating Temperature 5 C to 40 C 41 F to 104 F Storage Temperature 20 C to 60 C 4 F to 140 F Operating Humidity Max 80 up to 31 C 87 F decreasing linearly to 50 at 40 C 104 F Storage Humidity 80 Operating Altitude 2000meters 7000ft maximum Battery One 1 9V Battery NEDA 1604 Auto powe...

Отзывы: