EA11A-EU-EN v1.3 01/14
5
Specifications
General Specifications
Display
Multi-function LCD
Measurement Range
-50 to 1300
o
C (-58 to 1999
o
F)
Resolution
0.1
o
C/
o
F <200
o
, 1
o
C/
o
F >200
o
Data Memory
150 sets
Input Protection
20V maximum
Display Update Rate
2.5 times per second
Over range indication
"OL" appears on the LCD
Open input indication
"-------" appears on the LCD
Auto Power OFF
30 minutes (if no key is pressed)
Low battery indication
“BT” appears on the LCD
Power supply
6 AAA Batteries
Battery life
Approximately 110 hours with carbon zinc batteries
Operating Temperature
0 to 50
o
C (32 to 122
o
F)
Operating Humidity
< 80% RH
Storage Temperature
-10 to 60
o
C (14 to 140
o
F)
Storage Humidity
< 70% RH
Altitude
up to 2000 meters
Approvals
Dimensions
150x72x35mm (5.91x2.8x1.4")
Weight
Approx. 235g (8.29 oz.) with battery
Accuracy Specifications
Units
Range
Accuracy (@23 ± 5
o
C)
o
F
0
o
F to 1832
o
F
±(0.3% r 2
o
F)
-58
o
F to 0
o
F and
1832
o
F to 1999
o
F
±(0.5% r 2
o
F)
o
C
0
o
C to 1000
o
C
±(0.3% r 1
o
C)
-50
o
C to 0
o
C and
1000
o
C to 1300
o
C
±(0.5% r 1
o
C)
Temperature Coefficient
0.1 times the applicable accuracy specifications per
o
C from 0
o
C to
18
o
C and 28
o
C to 40
o
C (32
o
F to 64
o
F and 82
o
F to 104
o
F)
Note: Temperature accuracy does not include the accuracy of the type K probe.
Note: The temperature scale is based on the international temperature scale of 1990 (ITS90).
Copyright
©
2014 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in any form
ISO-9001 Certified
www.extech.com