background image

4                                      CO210-en-EU_V1.2   8/14 

ALARM 

The meter features an audible alarm which will give warnings when the CO

2

 concentration exceeds 

the set limit. There are two settable limits, an upper limit that initiates the alarm and a lower limit to 
stop the alarm. 
The meter emits beeps and displays the fan icon on the display when the CO

2

 level goes over the 

upper limit. The audible alarm can be stopped by pressing any key or it automatically stops when 
CO

2

 reading falls below the lower limit. 

If the beeper is temporarily shut down, it will sound again when readings fall below lower limit and 
then go over the upper limit again or it the user presses the RESET button for more than 1 second 
to activate it. 
The fan icon keeps flashing when beeps are manually shut down. It stops only when readings fall 
under the lower limit 

SETUP

(Alarm level, Temperature scale, Real Time clock)

Hold the SET button, in the normal mode, for more than 1 second to enter the setup mode. 

P1.

1

 CO

2

 ALARM: GOOD LEVEL 

When entering the setup mode, P1.

0

 and “CO2” are displayed on the LCD. Press the SET button 

again to go into P1.

1

 for setting the CO

2

 upper limit for the GOOD level. The current set value will

be blinking on the display. 

Press the LOG/ or MIN/MAX  button to increase or decrease the value. Each press adjusts in 
100ppm increments. The alarm range is 0 to 700ppm. 
When the value has been set, press the SET button to confirm the GOOD limit and to proceed to 
P1.

2

 for setting the upper NORMAL limit. Press the ESC button to exit without saving the setting. 

P1.

2

 CO

2

 ALARM: NORMAL LEVEL 

P1.

2

 is used for setting the CO

2

 upper limit for the NORMAL level. The current set value will be 

blinking on the display. 
Press the LOG/ or MIN/MAX  button to increase or decrease the value. Each press adjusts in 
100ppm increments. The alarm range is 700 to 1000ppm. 
When the value has been set, press the SET button to confirm the NORMAL limit and to proceed to 
P1.

3

 for setting the upper POOR limit. Press the ESC button to exit without saving the setting. 

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Содержание CO210

Страница 1: ...User s Guide CO2 Monitor and Datalogger Model CO210 www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com...

Страница 2: ...ncentration in ppm 2 Date and Time 3 Air Temperature 4 Relative Humidity Symbols ppm CO2 value GOOD CO2 air quality level NORMAL CO2 l air quality level POOR CO2 air quality level Air Temp Air Tempera...

Страница 3: ...clear the minimum and maximum value from memory DATA LOGGING The meter can record readings of CO2 temperature and humidity for long time environment monitoring The memory capacity is 15999 points 5 33...

Страница 4: ...tup mode P1 1 CO2 ALARM GOOD LEVEL When entering the setup mode P1 0 and CO2 are displayed on the LCD Press the SET button again to go into P1 1 for setting the CO2 upper limit for the GOOD level The...

Страница 5: ...onfirm the setting or press the ESC button to exit without saving and return to P2 0 P3 0 REAL TIME CLOCK Press the LOG button in P1 0 twice to access P3 0 for setting up real time clock Press the SET...

Страница 6: ...er Default P1 1 700ppm P1 2 1000ppm P1 3 1000ppm P2 1 C P4 1 No P5 0 SAMPLING RATE Press the LOG button in P1 0 three times to access P5 0 to set the logging sample rate The available range is 1 secon...

Страница 7: ...a sunny day with good ventilation and fresh air where the CO2 level is approximately 400 ppm Do not calibrate on a rainy day because high humidity will affect the CO2 level in air Do not calibrate in...

Страница 8: ...r 90 Display LCD Sensor Type CO2 NDIR non dispersive infrared technology Humidity Capacitance sensor Temperature air Thermistor Response CO2 2min for 90 step change Temp 2min for 90 step change RH 10m...

Страница 9: ...If above two methods faile return for repair E17 ABC mode of CO2 senor has failed and might cause wrong CO2 readings Send back for repair Temp Display E02 Air temp measurement is under the lower limit...

Страница 10: ...n resulting in permanent brain damage coma and even death Regulatory exposure limits ASHRAE Standard 62 1989 1000ppm CO2 concentration in occupied building should not exceed 1000ppm OSHA 5000ppm Time...

Отзывы: