4                                     42520es       Ver. 1.7    8/00

 

6.2 High InfraRed Energy Disturbances

 

Objects having low emissivity and objects with low temperatures but high emissivity emit 

little infrared energy. Because of this, measurements are adversely effected by powerful 

infrared energy radiated from nearby objects having high emissivity or high temperature. 

For example, when such objects are measured in sunlight, erratic measurements are 

caused by powerful radiated energy from the sun reflected on the surface of an object. 

 

6.3 Measurement Surface Hindrances

 

If the surface of an object to be measured is covered with frost, dust or other materials, 

surface cleaning must be performed before accurate measurements can be made. 

 

6.4

 

High Reflective Surfaces

 

If the surface of an object to be measured is highly reflective, apply masking tape or apply 

flat black paint to the surface since this paint has a known emissivity value. 

 

7. BATTERY REPLACEMENT 

A flashing display indicates that the battery voltage has fallen into the critical region (6.5 to 

7.5 V). Reliable readings can be obtained for several hours however, after the first 

appearance of the low battery indication. Open the Battery compartment, replace the 

battery, and close the compartment cover. 

 

 

8. CALIBRATION / REPAIR SERVICES 

Extech offers complete repair and calibration services for all of the products we sell. For 

periodic calibration, NIST certification or repair of any Extech product, call customer 

service for details on services available. Extech recommends that calibration be performed 

on an annual basis to insure calibration integrity. 

 

9. WARRANTY 

EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and 

workmanship for one year from date of shipment (a six month limited warranty applies on sensors and 

cables). If it should become necessary to return the instrument for service during or beyond the warranty 

period, contact the Customer Service Department at (781) 890-7440 for authorization. A Return 

Authorization (RA) number must be issued before any product is returned to Extech. The sender is 

responsible for shipping charges, freight, insurance and proper packaging to prevent damage in transit. 

This warranty does not apply to defects resulting from action of the user such as misuse, improper wiring, 

operation outside of specification, improper maintenance or repair, or unauthorized modification. Extech 

specifically disclaims any implied warranties or merchantability or fitness for a specific purpose and will not 

be liable for any direct, indirect, incidental or consequential damages. Extech's total liability is limited to 

repair or replacement of the product. 

The warranty set forth above is inclusive and no other warranty, whether written or oral, is expressed or 

implied. 

Copyright 

©

 

1999 Extech Instruments Corporation. All rights reserved 

including the right of reproduction in whole or in part in any form.

 

 

 

 

Содержание 42520

Страница 1: ...purchase of Extech s 42520 IR Thermometer This professional meter with proper care will provide years of safe reliable service 2 SAFETY Use extreme caution when the laser beam is ON Do not let the beam enter your eye another person s eye or the eye of an animal Be careful not to let the beam from a reflective surface strike your eye Do not use the laser near explosive gases or in other potentially...

Страница 2: ... 6 170 x 44 x 40 mm 3 2 Range Specifications Resolution Ranges 1o C F 0 to 600o F 20 to 315o C Accuracy 2 rdg or 4o F 2o C whichever is greater Accuracy notes 1 Accuracy specified for ambient temperature 2 Accuracy specified for emissivity of 0 95 Emissivity settings 0 95 fixed Distance Factor D S Approx 6 1 D distance S spot Measurement Field and Target Size Refer to Chart on top of meter or Fig ...

Страница 3: ...old the ON HOLD key and then press the BACKLITE key Repeat this procedure to turn the backlight OFF 5 6 Laser Pointer To turn the laser pointer ON press the LASER key while pressing and holding the ON HOLD key Repeat the procedure to turn the laser OFF 5 7 Measurement Considerations a Holding the meter by its handle point the IR Sensor toward the object under test b The object under test should be...

Страница 4: ...ll For periodic calibration NIST certification or repair of any Extech product call customer service for details on services available Extech recommends that calibration be performed on an annual basis to insure calibration integrity 9 WARRANTY EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for one year from date of shipment a six month limit...

Отзывы: