13                                     382860       Ver. 1.5    2/01

 

 

 

Main LCD

 

Left sub-LCD

 

Center sub-LCD

 

Right sub-LCD

 

Mode 0

 

Watts

 

AC Load 

Current

 

AC Voltage

 

Phase angle (cosine)

 

Mode 1

 

Watts

 

Rate Cost/hour

 

Accumulated 

Watts

 

Accumulated consumption 

cost in $

 

Mode 2

 

Accumulated 

Watts

 

EDIT the Rate 

Cost/hour 

 

Unit 1KW/Hr

 

NO DISPLAY

 

TABLE 1: Three Power Measurement Display Modes 

 

7.2.1 Explanation of terms in TABLE 1 

• 

Watts:

 True Power reading of load/appliance measurement 

• 

AC Load Current

: Current measurement for load/appliance under test 

• 

AC Voltage:

 Voltage measurement of load/appliance under test 

• 

Phase angle:

 Phase difference (from –90 to +90 degrees) between input 

source and load 

• 

EDIT the Rate Cost per hour in $

: This is where the user programs the Cost 

per hour (instructions on how to set this value are given in Section: 7.3)  

• 

Accumulated Consumption Cost in $:

 This is the total cost of consumption 

monitored since measurement session began (displayed in US$)

 

• 

Accumulated Watts: 

Wattage accumulated from when measurement session 

began up to the present (displayed in KW/H).

 

• 

KW/H: 

Kilowatt hours 

 

• 

Unit 1 KH/H: 

Reference value of 1KW/H

 

 

 

7.3 Programming Cost per Hour

 

1. Go to mode 2 using the UP/DOWN keys to scroll through the modes. The mode 

designator is the digit to the upper left of the left-most sub-LCD. 

2. The left-most sub-LCD is where the Cost per Hour is programmed. 

3. Press the SET/RESET key to begin editing the left most digit will flash. Enter the 

desired value via the UP/DOWN keys. While programming, to move to the next 

digit for editing, use the SET/RESET key. Digits are selected and ready for 

editing when they are flashing. 

4. Press the SET/RESET key until the digits cease flashing when finished editing 

the cost per hour value and then measure power in the desired mode 

Note

: Based on the consumption cost per hour programmed, the meter displays 

the expected rate cost per hour automatically. Currency is displayed in US dollars. 

 
7.4 Storing and Clearing Accumulated Measurement Data

 

The meter memorizes accumulated reading data when the POWER position of the 

Rotary Switch has been changed to another function after Power measurements 

have been made (this activates the memory utility). To recall accumulated data, re-

select POWER on the Rotary Switch, the meter displays stored data in EEPROM. 

To clear accumulated readings: Set the Cost per KW/H to 0000.0 in mode 2 (left 

sub-LCD display). 

 

7.5 General Power Measurement Notes:

 

1.  When POWER is selected via the Rotary Switch, automatic power-down 

works only when the watt measurement reading is zero. 

2.  When POWER is selected on the Rotary Switch, the FUNCTION key will not 

operate when pressed. 

3.  To save (store) the cost per hour, KW/H, and accumulated cost data, press 

the FUNCTION key or turn the Rotary Switch to another range AFTER 

DISCONNECTING THE POWER ADAPTOR. Never change the Rotary 

Содержание 382860

Страница 1: ...the correct measuring range to achieve best accuracy REL Relative function allows the operator to measure a value with respect to a preset reference value CMP Comparator function allows a comparative measurement to be made the upper lower limits can be defined MEM Memory and RCL Recall functions permit 10 measurements to be stored in memory for later recall AUTO POWER OFF extends the life of the b...

Страница 2: ...ons NOTES Accuracies are of reading and specified at 23o C with RH 75 for a period of 1 year after production Warm up time is 1 minute Mode Measuring Range Accuracy Resolution 400 mV 100 µV 4 V 1 mV 40 V 10 mV 400 V 0 3 1dgt 100 mV DC Voltage 1000 V 0 5 1dgt 1 V 400 µA 0 1µA 4mA 1 0 1dgt 1 µA 40mA 10 µA 400mA 0 8 1dgt 100µA 4A 1mA DC Current 20A 1 5 5dgt 10mA Accuracy Mode Measuring Range 40Hz 1KH...

Страница 3: ...0 Ohm Audible tone approx 40 Ohm Mode Measuring Range Accuracy Power Measurements AC Current AC Voltage True Power True Power Phase angle cosine To 16 A 91 to 120V To 100 W To 4000 W 90 to 90 degrees 2 5dgts 3 3dgts 5 10dgts 2 5dgts Power Factor 00 00 to 1 00 2 3 Overload Protection 400mA AC DC in the mA range Resistance Measurement 40 Mohm overload Protection 250 V DC AC rms Frequency Measurement...

Страница 4: ...ned voltage carrying components may be accessible Terminals can also carry voltage If it is necessary to open the unit for adjustment maintenance repairing or exchange of parts or modules the measuring unit has to be removed from all voltage sources Capacitors can still carry voltage even if the unit has been removed from all voltage sources Please make sure to use new fuses of the proper rating D...

Страница 5: ...t types The supplied test leads are rated for a maximum of 1200V The maximum rating of the multimeter is 1000 VDC max or 750 VAC rms Use special caution when working with voltage above 25 V AC and above 35 V DC SAFETY SYMBOLS This symbol adjacent to another symbol terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal i...

Страница 6: ...ity check modes UP DOWN Keys Used to 1 Enter reference values in the REL or CMP modes 2 Address stored values in the MEM and RCL modes 3 Manually set the measurement range in the R H function mode 4 Scroll through advanced Functions list UP forward DOWN backward BACKLIGHT ADJUST Key circular blue colored button Press this key to deepen the LCD backlight effect The backlight extinguishes after a 15...

Страница 7: ...ing frequency measurement To return from this function mode press the SET RESET key Auto Hold A H In this function the meter will automatically record the MIN MAX AVG readings taken since this mode was accessed These values are continuously updated To access A H press the FUNCTION key three times Then press the SET RESET key once Use the UP and DOWN keys to select the decimal location for the MIN ...

Страница 8: ...sub display the 2nd recalled value ref No 1 appears on the center sub display and the 3rd recalled value ref No 2 appears on the right side sub display and so on for the remaining locations CMP Comparison Comparison mode allows readings to be compared to user programmable High Low setpoints Press the FUNCTION key and then the DOWN key Press the SET RESET key once and the word LOW should begin flas...

Страница 9: ...c and read the current on the LCD During AC current measurements the middle sub display will show the frequency of the measurement and the main display indicates the current measurement 5 During DC current measurements the left most sub display is updated one second later than the main LCD The middle sub display two seconds later and the right sub LCD three seconds later Attention Never measure cu...

Страница 10: ... or 0 7 V Silicon if the diode is GOOD 4 Reverse the probe tips Red to the cathode and Black to the anode to check the diode s reverse status 5 If OL overload is displayed the diode is open or measures above 2 0 V If a value between 0 V and approx 2 0 is displayed then a good forward voltage drop exists 6 During diode tests the center sub display indicates GOOD if the diode is o k If the diode is ...

Страница 11: ...isplay 2 seconds later and the right 3 seconds later Important Transistor measurement notes Observe the sequence of connections E B C for the transistor Some Darlingtons or special power transistors contain internal base to emitter resistances that could cause erroneous measuring results The hFE measuring value is not an absolute measurement It only indicates if the transistor is operating or not ...

Страница 12: ...e right Always connect the correct contacts for all plug connections The labeling on the Power Adaptor matches the meter s labeled input terminals for convenience 3 Connect the appliance load under test not powered to the socket of the Power Adaptor 4 Power the load under test by inserting the three prong plug into a 117VAC outlet Readings can now be observed on the LCDS 5 The maximum AC current t...

Страница 13: ...per Hour is programmed 3 Press the SET RESET key to begin editing the left most digit will flash Enter the desired value via the UP DOWN keys While programming to move to the next digit for editing use the SET RESET key Digits are selected and ready for editing when they are flashing 4 Press the SET RESET key until the digits cease flashing when finished editing the cost per hour value and then me...

Страница 14: ... numbers 0 010 KHz 10Hz reference no 0 0 050 KHz 50Hz reference no 1 0 060 KHz 60Hz reference no 2 0 100 KHz 100Hz reference no 3 0 400 KHz 400Hz reference no 4 1 010 KHz 1010Hz reference no 5 2 021KHz 2021Hz reference no 6 4 042KHz 4042Hz reference no 7 8 084KHz 8084Hz reference no 8 10 24kHZ 10240Hz reference no 9 Attention Never short the output of the signal generator The output terminal and t...

Страница 15: ...SCOPE VIEW icon Follow the on screen help for specific operation instruction Also refer to the README file in the diskette s SCOPE subdirectory for operation hints 10 MAINTENANCE 10 1 Fuse Replacement Remove power to the meter and removal all connected cables and probes Use a suitable Phillips screwdriver to carefully open the case Remove the defective fuse s and replace it with a new one of the s...

Страница 16: ... 890 7440 for authorization A Return Authorization RA number must be issued before any product is returned to Extech Thesenderis responsible for shipping charges freight insurance and proper packaging to prevent damage in transit This warranty does not apply to defects resulting from action of the user such as misuse improper wiring operation outside of specification improper maintenance or repair...

Отзывы: