background image

3                           382200 and 382202 V3.1   07/06 

Operation 

CAUTION:

 Verify that the AC voltage setting matches your power source 

BEFORE applying power to the instruments. Do not connect to a voltage 
source that is greater. 

CAUTION

: Limit full load operation (>25watts for the 382200 and >45watts for 

the 382202) to 15 minutes continuous operation. Always allow a cool down 
period after full load operation. 

 

1. 

The Power Supply must be powered using the rated line voltage 10%. 

2. 

Before turning the power supply ON, remove any load and set the Voltage and 
Current Adjustment Knobs to their full counter-clockwise positions. 

3. 

Use the Current and Voltage Adjustment knobs for setting variable Current and 
Voltage output signals respectively. 

4. 

The LCD displays will indicate actual current and voltage outputs. 

5. 

Keep the meter's cooling vents (top and sides) clear of obstacles to prevent 
overheating. 

Current Limit 

To avoid having the load draw too much current, a current limit can be set as follows: 
With the voltage and current output adjustments set fully off, carefully short the positive 
and negative output terminals together. Turn the Output Voltage knob clockwise to the 
25% position (approx). Adjust the Output Current knob clockwise until the desired 
current level is displayed. Turn the Output Voltage knob counter-clockwise (current 
display will go to 0.00). Connect a load to the output terminals and adjust the voltage to 
the desired level. 

 

Warranty 

EXTECH INSTRUMENTS CORPORATION warrants the basic instrument to be free of 
defects in parts and workmanship for one year from date of shipment (a six month limited 
warranty applies on sensors and cables). If it should become necessary to return the 
instrument for service during or beyond the warranty period, contact the Customer Service 
Department at (781) 890-7440 EXTENSION 210 for authorization or visit 

www.extech.com

 

for more information. 

A Return Authorization (RA) number must be issued before any 

product is returned to Extech.

 The sender is responsible for shipping charges, freight, 

insurance and proper packaging to prevent damage in transit. This warranty does not 
apply to defects resulting from action of the user such as misuse, improper wiring, 
operation outside of specification, improper maintenance or repair, or unauthorized 
modification. Extech specifically disclaims any implied warranties or merchantability or 
fitness for a specific purpose and will not be liable for any direct, indirect, incidental or 
consequential damages. Extech's total liability is limited to repair or replacement of the 
product. The warranty set forth above is inclusive and no other warranty, whether written or 
oral, is expressed or implied. 

Calibration and Repair Services 

Extech offers repair and calibration services for the products we sell. Extech also provides 
NIST certification for most products. Call the Customer Service Department for information 
on calibration services available for this product. Extech recommends that annual 
calibrations be performed to verify meter performance and accuracy. 