Total Harmonic Distortion
(THD-F with respect to the 50 or 60 Hz, min. value at the 50 or 60 Hz > 80V
and > 20 A, 1 to 50
th
Harmonics. If the voltage or current is 0 at 50 or 60 Hz, all of the percentage (%)
displays are 0)
1000A Clamp
Range Resolution
Accuracy
0.0 to 20%
0.1%
± 2%
20 to 100%
01%
± 6% of reading ± 1%
100 to 999.9 %
0.1%
± 10% of reading ± 1%
Total Harmonic Distortion
(THD-F with respect to the fundamental frequency, min. value at 50 or 60 Hz is
greater than AC 80V, greater than 10% of the range. Calculation is done over 1 to 50
th
Harmonics. If the
voltage or current is 0 at 50 or 60 Hz, all the percentage (%) displays are 0).
100A Clamp
Range Resolution
Accuracy
0.0 to 20.0 %
0.1%
± 1%
20.0 to 100%
0.1%
±3% of reading ± 5%
100 to 999.9%
0.1%
±10% of reading ±10%
3000A Flexible Clamp
Range Resolution
Accuracy
0.0 to 20%
0.1%
± 2% of range
20 to 100%
0.1%
±6% of range ± 1%
100 to 999.9%
0.1%
±10% of range ±1%
Peak Value of AC Voltage (peak value > 5V) or AC Current (peak value > 20A), VT=1
Model 1000A Clamp
Range
Sampling Time
Accuracy of Reading
50 Hz
39 µs
± 5% ± 30 digits
60 Hz
33 µs
± 5% ± 30 digits
Peak Value of AC Voltage (peak value > 5V) or AC Current (peak value > 10% of the range), VT=1
Model 100A Clamp
Range
Sampling Time
Accuracy of Reading
50 Hz
19µs
± 5% ± 30 digits
60 Hz
16µs
± 5% ± 30 digits