382075 V2.2V 9/10
17
AC/DC Current (True RMS, Crest Factor < 4):
Accuracy (of rdg)
Overload
Protection
Range
Resolution
DC, 50 / 60 Hz
40 - 400Hz
0 - 200A
0.1A
±1.5% ±5dgts
±2.0% ±5dgts
AC 3000A
200 - 500A
0.1A
±2.0% ±5dgts
±2.5% ±5dgts
500-2000A
1A
±2.5% ±5dgts
±3.0% ±5dgts
AC/DC KVA (Apparent Power)
KVA = V*A / 1000
AC KVAR (Reactive Power, PF < 0.91)
Range Resolution
Accuracy
Remark
0 to
1200KVAR
0.1
1 KVAR or
+/-3%rdg+9d
(whichever is greater)
For more information, refer to the
section entitled "DC/AC 1
φ
2W
Apparent/Reactive Power
measurements"
Phase Angle (User must zero the current reading before taking measurements)
Range Accuracy
Sensitivity
Remark
-90
°
to +90
°
±
2.0
°
V > 100V, A > 10A
Zero Crossing Detection
Note: If current signal is not detected, the phase angle will be left blank in LCD.
Frequency (if < 10 Hz, then the Hz Display = 0)
Range Accuracy
Sensitivity
50/60 Hz
±2 dgts
V > 1V, A > 5A
10 - 400 Hz
0.5% ± 2dgts
Power Factor (PF)
Range Max.
Resolution
PF = KW / KVA
0.2 to 1
0.01