background image

380820-en-GB_V1.5   3/15 

13. 

$: 

Calculates the power rate. Type the power rate first, then press $ to confirm. When the unit 

is in OPER mode, the power rate per day/month/year will be displayed. 

14. 

OCP: 

Preset value of the Over Current Protection. Type the value for over-current protection 

and then press OCP to confirm. 

15. 

Hz:

 Press to enter the frequency programming mode for the front output. Type the desired 

frequency and then press Hz to confirm. 

16. 

FUNC: 

Press to view

 

THD (total harmonic distortion), voltage, current, and harmonics (1st to 

50th) for the output power. 

17. 

STORE:

 Press to save the currently programmed voltage and frequency to one of the memory 

locations (1 to 99) and press V.ENTER to confirm. The currently selected voltage and 
frequency will be stored in the selected memory location. 

 
 
 
 

1. 

Ventilation: 

Please keep these openings clean and always leave room for ventilation at the 

rear of the unit. 

2. 

POWER INPUT:

 Socket for external power input (contains a 10A fuse). The input is selectable 

(110V or 220V) and must match the voltage selected by the Voltage Switch. 

3. 

RS232 Connector: 

For

 

connection to a PC serial port. 

4. 

Voltage Switch: 

Selects the voltage (110 or 220V) of the input power. 

5. 

Rear Output Socket 

CAUTION! The rear output socket is “LIVE” whenever the instrument power 
cord is plugged in even when the power supply is shut off.  

 

 

1

2

3

4

5

!

Содержание 380820

Страница 1: ...User s Guide Universal AC Power Source AC Power Analyzer Model 380820...

Страница 2: ...ly to a mains outlet with an earth ground connection Do not place the equipment on damp or wet surfaces Do not subject the equipment to direct sunlight or extreme temperatures Do not subject the equip...

Страница 3: ...r loss of life and prevent damage to this product Earth Ground Terminal This unit has been tested and meets or exceeds the following standards 1 EN61326 1 2006 CISPR11 IEC EN 61000 3 2 2006 IEC EN 610...

Страница 4: ...ft 4 LCD Liquid Crystal Display 5 RECALL Recalls stored voltage and frequency values there are 99 sets of preset values in the memory The currently selected voltage and frequency values will clear whe...

Страница 5: ...s to save the currently programmed voltage and frequency to one of the memory locations 1 to 99 and press V ENTER to confirm The currently selected voltage and frequency will be stored in the selected...

Страница 6: ...ic Test Setup Diagram Connect the Unit Under Test to the 380820 Front and Rear Output Select When the unit is switched ON use the OUTPUT button to select the display for the Front or Rear output data...

Страница 7: ...t is 0V however the output terminals are still physically connected to the internal circuit Frequency Output Front Output Rear Measurement 1 Type the desired frequency value and then press the Hz butt...

Страница 8: ...t offers 99 memory locations To Store 1 Ensure that the desired voltage and frequency values are correctly set 2 Press the STORE button verify STORE _ appears on display 3 Assign a memory location num...

Страница 9: ...press the button 2 When the unit is in OPER Output Mode changing the OCP value or the Voltage value will switch the mode from OPER to STBY standby Therefore pay careful attention to this operation 3 I...

Страница 10: ...the FUNC button is pressed the LCD will display Voltage THD and harmonics n 1 50 Use the buttons to select the harmonics to be displayed 3 When the FUNC button is pressed again the LCD will display t...

Страница 11: ...the output will be placed in STANDBY mode and the LCD will display OVERHEATED A displayed timer indicates the remaining cooling time that must elapse before the unit can be switched ON again When the...

Страница 12: ...2 minutes if the output is less than 100VA If the output is less than 25VA then the output can run continuously Rear Panel Specifications Input power is looped through to the REAR output Max 2000VA 2...

Страница 13: ...ity 85 RH Operation Temperature 4C to 60C 40 F to 140 F Accessories Power cord Software CD ROM RS232C to USB cable Dimensions 260mm x 151mm x 305 mm 10 2 x 5 9 x 12 0 Weight 9 9 kg s 21 8 lb Copyright...