Maintenance
MaxTester
409
Recalibrating the Unit
Recalibrating the Unit
EXFO manufacturing and service center calibrations are based on the
ISO/IEC 17025 standard (
General Requirements for the Competence of
Testing and Calibration Laboratories
). This standard states that calibration
documents must not contain a calibration interval and that the user is
responsible for determining the re-calibration date according to the actual
use of the instrument.
The validity of specifications depends on operating conditions. For
example, the calibration validity period can be longer or shorter depending
on the intensity of use, environmental conditions and unit maintenance, as
well as the specific requirements for your application. All of these elements
must be taken into consideration when determining the appropriate
calibration interval of this particular EXFO unit.
Under normal use, the recommended interval for your unit is: 2 years.
For newly delivered units, EXFO has determined that the storage of this
product for up to six months between calibration and shipment does not
affect its performance (EXFO Policy PL-03).
Содержание MAX-800 Series
Страница 1: ...www EXFO com User Guide MAX 800 Series MaxTester Application ETHERNET AND TRANSPORT TESTER...
Страница 16: ......
Страница 18: ......
Страница 26: ......
Страница 43: ...Test Setup Test Applications MaxTester 35 SONET SDH BERT For SDH BERT...
Страница 48: ...Test Setup Test Applications 40 MAX 800 Series SONET SDH DSn PDH BERT For SDH...
Страница 62: ......
Страница 208: ......
Страница 414: ......
Страница 426: ......
Страница 446: ...Glossary 438 MAX 800 Series Ethernet Cables Straight Through Cable 1000 Mbit s Crossover Cable 1000 Mbit s...
Страница 473: ...Glossary MaxTester 465 SONET DSn SDH PDH The GUI Grid indicates the TUG 2 x and TUG 3 x values...
Страница 476: ......
Страница 492: ...Index 484 MAX 800 Series Z Z0 344 Z1 346 Z2 349 Z3 353 Z4 353 Z5 353 Z6 355 Z7 356 Zero Code Suppression 169...
Страница 494: ...0 5 1 5 48 5 0 176 W d y K y K D...