Preparing for Automation
140
IQS-600
Getting Optimum Performance from Your Integrated Qualification System
For information on cabling, see
Preparing Hardware for GPIB Control
Bus timing (GPIB control only)
: Required if you intend to work with
IEEE 488.1, for it affects the handshake rate of this protocol. It indicates
the minimum amount of time a GPIB controller waits, once data is
placed on the bus, before setting the data valid bus line (DAV) to true.
The possible bus timings are:
If your setup supports it, select the very-high-speed timing to get the
best performance.
Output unit (GPIB, RS-232, ActiveX and TCP/IP):
The IQS controller
unitcan return results in
linear units (for example, watts)
log units (for example, dBm)
Since internal units are linear, you will get optimal performance by
using linear units for output (no need for an internal conversion to log).
Note:
You must make the choice of output unit for each instrument offering such
a feature. Refer to the user guide of each optical instrument for a list of
available commands and queries.
Mode
Bus timing delay
Normal
1100 ns
High speed
500 ns
Very high speed
350 ns
Содержание IQS-600 Series
Страница 1: ...User Guide Integrated Qualification System IQS 600...
Страница 22: ......
Страница 28: ......
Страница 106: ......
Страница 132: ......
Страница 194: ...Using IQS Products in an Automated Test Environment 186 IQS 600 Standard Status Data Structure...
Страница 252: ......
Страница 274: ......
Страница 336: ......
Страница 366: ...Using the IQS 9600 Utility Module 358 IQS 600...
Страница 367: ...Using the IQS 9600 Utility Module Integrated Qualification System 359...
Страница 368: ......