Preparing for Automation
136
IQS-600
Preparing Hardware for GPIB Control
Preparing Hardware for GPIB Control
If you intend to use GPIB to remotely control your instruments, your
controller must have been equipped with the optional GPIB card.
To ensure the optimum efficiency of your system, EXFO recommends that
you follow these restrictions:
For the IEEE 488.1 protocol:
Maximum of 15 devices physically connected to each GPIB bus.
Maximum separation of 4 m between two devices and an average
separation of 2 m over the entire range (bus).
Maximum total cable length used in the system is 20 m.
At least two-thirds of the connected devices must be turned on.
For the HS488 protocol:
Maximum of 15 devices physically connected to each GPIB bus.
Maximum total cable length used in the system is 15 m (that is, 1 m per
device physically connected to the GPIB bus).
All connected devices must be turned on.
For each meter of cable used in the system, there must be a device
connected to it or an equivalent device load.
Содержание IQS-600 Series
Страница 1: ...User Guide Integrated Qualification System IQS 600...
Страница 22: ......
Страница 28: ......
Страница 106: ......
Страница 132: ......
Страница 194: ...Using IQS Products in an Automated Test Environment 186 IQS 600 Standard Status Data Structure...
Страница 252: ......
Страница 274: ......
Страница 336: ......
Страница 366: ...Using the IQS 9600 Utility Module 358 IQS 600...
Страница 367: ...Using the IQS 9600 Utility Module Integrated Qualification System 359...
Страница 368: ......