SCPI Command Reference
66
IQS-2150
Product-Specific Commands—Description
:SOURce[1..n]:POWer:ATTenuation?
Description
This query returns a value corresponding to the
internal power attenuation of the source.
*RST sets the attenuation to 0 dB.
Syntax
:SOURce[1..n]:POWer:ATTenuation?[<wsp>MA
Ximum|MINimum|DEFault]
Parameter(s)
Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax
<Attenuation>