Maintenance
CD/PMD Analyzer Source
41
Upgrading the Embedded Software
7.
When a message about waiting for a evice handshake appears, turn on
the CD/PMD Analyzer Source.
The unit display remains off, the unit beeps once and the upgrade
program starts automatically. A progress bar on the computer screen
indicates the upgrade status. Once the software upgrade is complete,
the message about restarting the unit appears.
8.
If the software upgrade was performed under Windows 2000, an error
message to the effect that the
LO0006 NTVDM has encountered a
system error and to select close to terminate the application
is
displayed. Click
Close
to hide the dialog box.
9.
Turn the CD/PMD Analyzer Source off, and then on again.
Some units will display the new version number at startup, otherwise
press the up and right arrow keys together while the unit is turned on.