Controlling the Source Remotely
24
FLS-5834A
Setting the GPIB Address
Setting the GPIB Address
If GPIB is selected as the remote command mode, you can select the GPIB
address you want to use from 1 to 30 (default value is 12).
To set a GPIB address:
1.
Press the Setup key.
2.
Use the up/down or left/right arrow keys to select
GPIB Addr.
The
current GPIB address is displayed.
Note:
If you are in RS-232 mode, the GPIB address cell will display
N.A
.
You must
change the communication mode to
GPIB
before setting an address.
3.
Press ENTER, then use the up/down arrow keys to select a GPIB
address between 1 and 30.
4.
Press ENTER to confirm your choice.
Refresh Rate
Backlight
Contrast
Video Mode
Exit
8 Hz
ON
STD
RS232 / GPIB
GPIB Addr.
Baud Rate
Flow Ctrl
GPIB
12
N.A.
N.A.