Controlling the Source Remotely
CD/PMD Analyzer Source
23
Setting the Remote Command Mode
Setting the Remote Command Mode
To remotely control the CD/PMD Analyzer Source, you must set a GPIB
address or activate the RS-232 port.
To set a remote command mode:
1.
Press the Setup key.
2.
Use the up/down or left/right arrow keys to select
RS232 / GPIB
. The
current setting is displayed.
Note:
If
GPIB
is currently selected and you want to specify a GPIB address, see
Setting the GPIB Address
on page 24
.
3.
Press ENTER to access the
RS232 / GPIB
edit box.
4.
Use the up/down arrow keys to toggle between
GPIB
and
RS232
.
5.
Press ENTER to confirm.
If you selected
RS232
, the
GPIB Addr.
menu option is deactivated (
N.A
.
is
displayed).
If you selected
GPIB
, the
Baud Rate
and
Flow Ctrl
menu options are
disabled (
N.A
.
is displayed). If necessary, you can change the GPIB address.
Refresh Rate
Backlight
Contrast
Video Mode
Exit
8 Hz
ON
STD
RS232 / GPIB
GPIB Addr.
Baud Rate
Flow Ctrl
GPIB
12
N.A.
N.A.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание FLS-5834-EI
Страница 11: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 23: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 27: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 63: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...