Performing Measurement Scans
CT440
37
Configuring the CT440 for Measurements
3.
Set the wanted scanning options according to the instructions given in the following
table.
4.
Click the button or anywhere on the screen outside the window to exit.
Parameter
Description
Mode
Scanning mode of the CT440:
Single
The CT440 performs a scan of the wavelength range set in the
Start/Stop
fields and then stops.
On CT440 with PDL option: if a PDL measurement is required by one
or more traces (see
scan is composed of 4 or 6 successive sweeps (depending on the
selected
PDL Method
).
Continuous
The CT440 performs a continuous series of scans in accordance with
the interval set in the Interval parameter (see
Interval (s)
below),
until you click the
Abort
button.
Single Trig
The CT440 waits for an external trigger on the
Trigger In
BNC
connector to start a single scan. For more details, see
Continuous Trig
The CT440 waits for an external trigger on the
Trigger In
BNC
connector to start continuous scanning. For more details, see
PDL Method
CT440 with PDL option only.
4 States
: the CT440 performs 4 polarization-controlled successive
sweeps to achieve the PDL measurement.
6 States
: the CT440 performs 6 polarization-controlled successive
sweeps to achieve the PDL measurement. This provides the best
measurement performance but it takes longer.
Interval (s)
Applies to
Continuous
scan mode only.
Number of seconds between the beginning of two successive scans.
If the interval set is greater than the scan time, the CT440 waits
before the next scan.
If the period of time is lower than the scan time, the CT440
immediately performs the next scan.
In case of PDL measurement, the interval is observed between two
series of 4 or 6 successive sweeps.