25
Care and Maintenance
*Automatically applied when termperature is within normal body temperature range, otherwise reads
surface temperature.
Calibration Verification Procedure
All Exergen infrared thermometers are designed to permanently maintain their accuracy, and normally
recalibration is not required unless the thermometer has been physically damaged or experiences
component failure. In the unlikely event that recalibration may be required, the thermometer must be
returned to Exergen for the procedure.
However, calibration can be verified in the lab or clinical units quite easily using a device known as a
portable blackbody. A portable blackbody is a reference heat generator (see below), which is a self-
contained device providing a stable reference target temperature in the clinical temperature range.
The device can be used to verify the calibration of any Exergen thermometer in question, or for quality
checks done on a routine basis. The verifier operates with either a 9-volt power supply plugged directly
into any 120 vac wall receptacle, allowing extended use in the laboratory, or a 9-volt battery for
portable use on the nursing floors.
There are two ways to use the portable blackbody to verify the calibration accuracy of the thermometer
in question, either (1) with a certified master reference infrared thermometer, or (2) by using two
identical thermometers as a reference against the one in question.
• Specifications
LXTA/TAT-5000
LXTA/TAT-5000
LXTA/TAT-5000
LXTA/TAT-5000
TAT-4000
TAT-4000
TAT-4000
TAT-4000
Accuracy
± 0.2 F or 0.1 C
± 0.2 F or 0.1 C
Temperature Range
60 to 110 F (16 to 43 C)
61 to 110 F (16 to 43 C)
Arterial Heat Balance Range
for Body Temperature*
94 to 110 F (34.5 to 43 C)
94 to 110 F (34.5 to 43 C)
Operating Environment
60 to 110 F (16 to 43 C)
60 to 104 F (16 to 40 C)
Resolution
0.1 F or C
0.1 F or C
Response Time
~ 0.1 seconds
~ 0.04 seconds
Battery Life
~5,000 readings
~5,000 readings
Time Displayed on Screen
15 seconds
30 seconds
Size
2" x 8" x 1.25"
(5 cm x 20 cm x 3 cm)
2" x 8" x 1.25"
(5 cm x 20 cm x 3 cm)
Weight
8.3 oz (235 gm)
7.0 oz (199 gm)
Display Type and Size
Large bright LED's
Large bright LED's
Construction Method
•
•
•
• Industrial duty impact resistant
casing
•
•
•
• Hermetically sealed sensing
system
•
•
•
• Stainless steel probe
•
•
•
• Industrial duty impact resistant
casing
•
•
•
• Hermetically sealed sensing
system
•
•
•
• Aluminum probe
Warranty
Lifetime
Lifetime