Care and Maintenance
Instructions for Fahrenheit or Celsius Conversion
The Temporal
Scanner
can be used in either °F or °C. To convert from one scale to the other, the only tool necessary
is a paper clip or the tip of a small screwdriver.
FFFFFor °F/°C Con
or °F/°C Con
or °F/°C Con
or °F/°C Con
or °F/°C Convvvvver
er
er
er
ersion
sion
sion
sion
sion:
•
Loosen single screw on bottom of case and remove battery cover.
•
Lift battery out of the way.
TTTTTAAAAATTTTT-4000
-4000
-4000
-4000
-4000
•
Locate the little switch to the right of the battery as indicated in the
drawing, and with the tip of the paper clip or screwdriver, slide up or
down to the opposite position.
•
Remove the paper clip or screwdriver.
•
To return to the original setting, repeat the process.
LXT
LXT
LXT
LXT
LXTA/T
A/T
A/T
A/T
A/TAAAAATTTTT-5000:
-5000:
-5000:
-5000:
-5000:
•
Make sure the display is blank.
•
Insert the tip of the paper clip or screwdriver into the slot shown on the drawing, directly below and to the left
of the battery, and push in to activate.
•
While pressing the little switch, slide your finger around the other side of the instrument and turn the instru-
ment on by pressing the red button.
•
To return to the original setting, repeat the process.
D
D
D
D
D
ISPLA
ISPLA
ISPLA
ISPLA
ISPLAYYYYY
D
D
D
D
D
IA
IA
IA
IA
IAGNOS
GNOS
GNOS
GNOS
GNOSTICS
TICS
TICS
TICS
TICS
C
C
C
C
C
HAR
HAR
HAR
HAR
HARTTTTT
The following chart summarizes the fault conditions, and the associated indications.
C o n d itio n
C o n d itio n
C o n d itio n
C o n d itio n
Dis p la y
Dis p la y
Dis p la y
Dis p la y
So u n d
So u n d
So u n d
So u n d
LX TA / TA T-5 0 0 0
LX TA / TA T-5 0 0 0
LX TA / TA T-5 0 0 0
LX TA / TA T-5 0 0 0
TA T-4 0 0 0
TA T-4 0 0 0
TA T-4 0 0 0
TA T-4 0 0 0
High Target
HI
1 beep/second
>110 °F (43 °C)
>110 °F (43 °C)
Low Target
LO
1 beep/second
<60 °F (16 °C)
<61 °F (16 °C)
High Ambient
HI A
1 beep/second
>110 °F (43 °C)
>104 °F (40 °C)
Low Ambient
LO A
1 beep/second
<60 °F (16 °C)
<60 °F (16 °C)
Low Battery
N/A
2 beeps/second
bAtt
Dead Battery
– – – –
2 beeps/second
– – – –
Processing Error
E-## or Err
Restart. Return to Exergen for repair if error message persists.
TAT-4000
LXTA/TAT-5000