background image

 

 

 

 

Biological Indicator Incubator 

Operator’s Manual 

 

 

Revision 2.0

 

Содержание INC-13 Series

Страница 1: ...Biological Indicator Incubator Operator s Manual Revision 2 0...

Страница 2: ...ntained Biological Indicators SCBIs and Spore Ampoules Please read the manual carefully before using the incubator Please check the incubator and Packing List Page 13 to ensure all components have bee...

Страница 3: ...nstallation 4 Technical Data and Trouble Shooting 5 Safety Precautions 6 Operation Guide 7 Control Elements 7 Setting Incubation Time 7 Setting Temperature 7 Programming 8 Temperature Calibration 9 10...

Страница 4: ...ic Should the incubator prove defective Excelsior Scientific will exchange the defective device or refund the purchase price as Excelsior Scientific determines This warranty shall not apply if the inc...

Страница 5: ...as the most resistant to sterilisation processes The spores are a critical element to providing assurance that a sterilisation process was effective When incubated under the proper conditions the bact...

Страница 6: ...Incubator Diagram 3...

Страница 7: ...the main switch located on the back of the incubator The incubator is ready to operate when the display becomes visible See Setting the Temperature and Incubation Time to adjust these parameters prior...

Страница 8: ...Time 15 Minutes Ambient Temperature 5 C to 35 C Dimensions 120x152x112 mm Weight 0 85 kg Troubleshooting Display Cause Solution None No power Plug in incubator Check connections of adaptor and cord pl...

Страница 9: ...iorscientific com or Tel 44 7813762926 Use and store the incubator on a level horizontal surface in a location with little or low levels of dust moisture and sun Select a location with good air moveme...

Страница 10: ...Setting the Temperature The default temperature is 56 0 C when other temperatures are required press PROG and key simultaneously The following will be displayed SetTemp 56 0C Then press key to move th...

Страница 11: ...Setting Temperature to adjust temperature set point Press PROG key again to display incubation time After 3 seconds the display will return to temperature Press START STOP key to start timing The wil...

Страница 12: ...tment can be made as follows After turning on the power to the incubator make sure the temperature on the display is 35 C If above 35 C wait until the incubator cools to less than 35 C Place a thermom...

Страница 13: ...cleaned with alcohol to assure the heat is transferred effectively to the SCBI and to reduce the potential for contamination with microbial organisms other than the indicator organism post exposure D...

Страница 14: ...Incubator Wiring Diagram 12...

Страница 15: ...Packing List Item Type Quantity Incubator 13mm 10 well 1 Adaptor 24VC 2A 1 Plug Cord American European or United Kingdom 1 Transparent Cover Not Applicable 1 Operator s Manual Not Applicable 1 13...

Страница 16: ...Number Part Name AS27 00 05 Front Panel AS69 00 01 Heating Film 24V35W AS28 00B 20091008 Control Board N A DC Power Supply 48W 24V 2A AMT 130 Temperature Protection Switch BR 12C 11L On Off Switch HC1...

Страница 17: ...Distributed by Excelsior Scientific Ltd Excelsior House 3 Stow Gardens Wisbech PE13 2HS United Kingdom Tel 44 0 7813762926 sales excelsiorscientific com www excelsiorscientific com 15...