EXCELL PRECISION CO., LTD.
2-3-5 Noise Filter Setting
Select in the General Function setting mode
to set the noise filter
setting.
When modifying , the parameters of remain un-altered.
DISPLAY
ACTIONS
NOTE
Returning to the zero point setting
Use
ON/ZERO
or
UNITS
key to key in the
parameters or zero point
Default setting = 0
0
⇒
No skip 5
⇒
skip 5d
1
⇒
skip 1d 6
⇒
skip 6d
2
⇒
skip 2d 7
⇒
skip 7d
3
⇒
skip 3d 8
⇒
skip 8d
4
⇒
skip 4d 9
⇒
skip 9d
When the weight on the scale is over
1/3 full capacity, the function is on.
Digital switch & Stabilization range
setting
LCD displays the last parameter setting
Filter parameter setting
LCD displays the last parameter setting
Filter parameter setting
Use
ON/ZERO
or
UNITS
keys to key
in the parameters. Default setting = 5
Parameter 0 ~ 9, the larger the number,
the faster the filter response. Fast
response can lead to weight instability.
Digital switch & Stabilization range
setting
Use
ON/ZERO
or
UNITS
keys to key in
the parameters. Default setting = 0
Parameter 0 ~ 9, the larger the number
the more stable the weight.
Returning to zero point setting
LCD displays the last status
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key to complete
general setting procedures
Press
TARE
key
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Use
ON/ZERO
or
UNITS
key to key in the parameters
Press
TARE
key
Key in the parameter “00”
Setting is only
available for non-approval models.