![Excelitas Technologies X-Cite mini+ Скачать руководство пользователя страница 29](http://html1.mh-extra.com/html/excelitas-technologies/x-cite-mini/x-cite-mini_user-manual_2449447029.webp)
29
X-Cite® mini+ User Guide
035-00587R rev. 5
If an application requires lower power levels, strategies to avoid fluctuations include:
Increase power level, and reduce exposure time to compensate for brighter signal.
Increase power level, and use a neutral density filter or iris in the microscope light train to
reduce intensity to an appropriate level for the specimen.
9
Regulatory
9.1
Product Safety and Electromagnetic Compatibility
The X-Cite mini+ has been tested and found to comply with product safety and electromagnetic
compatibility requirements. For a complete list of tests and for certification details, please contact your
X-Cite representative or visit
https://www.excelitas.com/product/x-cite-mini-compact-led-illumination-
9.1.1
Optical Safety
The X-Cite mini+ is classified as Risk Group 3 according to IEC 62471: Photobiological Safety of Lamps
and Lamp Systems.
Resulting Classification and Labelling:
Hazard
All models
Actinic UV
Risk Group 3
Near UV
Risk Group 3
Blue Light
Risk Group 2
Retinal Thermal Weak Visual Exempt Group