i.

 

Select “Color” and scroll up /down through the options until desired color is on 
LCD. 

ii.

 

 Click dial to save selection and return to “LCD” menu. 

e.

 

Select “Exit” to go back to the main menu, or press and hold dial to return to the home 
screen.   

4.3.3

 

Favo – Favorite Intensity Setting  

In the FAVO sub menu, a favorite, commonly used intensity setting can be saved.  When this 
mode is enabled, the intensity level will instantly jump to this setting with a double-click of the 
dial.   A second double-click will toggle back to the previous intensity level.      

a.

 

Select “Favo” option from main menu. 

b.

 

To set/change the favorite intensity: 

i.

 

Select “Int” and scroll through % intensity settings until the desired level is 
reached.    

ii.

 

Click dial to save selection and return to “Favo” menu.   

iii.

 

TIP: Scrolling through % settings in this menu will not actually change the output 
in real time, even if LED is on.  The favorite intensity should be determined 
before entering this menu.   

c.

 

To enable/disable favorite intensity mode: 

i.

 

Select “Enbl” and scroll up/down to the desired setting.   

 

“On” will enable  (i.e. double-click speedDIAL = Favorite setting) 

 

“Off” will disable (i.e. double-click speedDIAL = do nothing)  

ii.

 

Click dial to save selection and return to “Favo” menu.   

d.

 

To adjust the delay between clicks in the “double-click”: 

i.

 

Select “Spd” and scroll up/down to the desired delay setting.  “1” will provide 
the shortest delay, “10” will be the longest.   

ii.

 

Click dial to save selection and return to “Favo” menu.   

iii.

 

TIP: If double-clicking results in the light being turned OFF instead of going to 
favorite intensity, increase the delay setting.    

iv.

 

Note: Setting long delay times will result in a longer response time for regular 
ON/OFF control with speedDIAL (the system must wait to see if the single-click 
turns into a double-click).  Response times for TTL, USB or foot pedal on/off 
control will not be affected. 

4.3.4

 

TTL –TTL Mode Control 

In the TTL menu, TTL mode can be enabled for experiments that require rapid LED ON/OFF 
control. For further information on TTL mode, refer to the External Control, TTL section 5.2.   
NOTE:  TTL mode MUST be enabled for the system to respond to a TTL signal

a.

 

Select “TTL” option from main menu. 

b.

 

To enable/disable TTL mode: 

i.

 

Select the “Enbl” menu item. 

ii.

 

Scroll to the desired setting, “On” to enable TTL or “Off” to disable TTL.  

Содержание X-Cite mini+

Страница 1: ...X Cite mini User Guide Applies to X Cite mini Series Models XMPS XMPL XTMS XTML XT120Lm X Cite 120LEDmini...

Страница 2: ...ans without the prior written consent of Excelitas Canada Inc Every effort has been made to ensure information in this manual is accurate however information in this manual is subject to change withou...

Страница 3: ...3 7 Srvc Service Data 16 5 Operation External Control 17 5 1 USB RS 232 17 5 1 1 Driver Installation via internet 17 5 1 2 Driver Installation via ZIP file 18 5 1 3 Verify Installation Get COM Port Nu...

Страница 4: ...5 China RoHS 31 9 6 Australia RCM 31 10 Warranty Repairs 32 10 1 Warranty Terms 32 10 2 Returning equipment to Excelitas Technologies 32 11 Contact Information 33 11 1 General 33 11 2 Technical Suppor...

Страница 5: ...or use in fluorescence microscopy illumination It allows researchers the ability to excite fluorescence in samples being studied and characterize their location or behavior Excelitas Technologies Corp...

Страница 6: ...tting end of the LED head The light could severely damage the cornea and retina of the eye if the light is observed directly Eye shielding must be used at all times as well as clothing to protect expo...

Страница 7: ...arm function must be provided by the user to indicate a malfunction in the associated equipment used 9 Should this X Cite unit be used in a manner not specified by Excelitas Technologies the protectio...

Страница 8: ...uction Sheet User Guide Software Driver downloads are available https www excelitas com product x cite mini compact led illumination system b Hex Tool 3mm c USB Cable d Power Cord e Safety Precautions...

Страница 9: ...t and accessories from the shipping carton b When removing the LED Head and miniCUBE from the carton ensure that both components are supported and there is always some slack in the cable Excessive str...

Страница 10: ...ee 3 fasteners provided The two 2 dowel pins on the LED Head will allow you to install the Flange quickly and in the proper position as required by your microscope Figure 4 Installing Microscope Flang...

Страница 11: ...the pin on the microscope fitting hold the LED head firmly against the light port and twist the collar to lock the components together iv Olympus Tighten the hex fastener s with the 3mm Hex Tool Depe...

Страница 12: ...ord supplied or one with an equivalent rating 4 Operation Manual Control 4 1 The Basics 1 Start the unit a Flip the rocker switch on the front of miniCUBE to turn X Cite mini on b The system will have...

Страница 13: ...tly selected menu option click the dial to make a selection To adjust settings turn the dial To exit the setting adjustment click the dial To exit menu system at any time press and hold the dial for o...

Страница 14: ...scroll through the time out options which are Off and 1 thru 999 seconds in 1 second increments Off will turn off the backlight timer i e backlight will always be on ii Click dial to save selection a...

Страница 15: ...o the desired setting On will enable i e double click speedDIAL Favorite setting Off will disable i e double click speedDIAL do nothing ii Click dial to save selection and return to Favo menu d To adj...

Страница 16: ...e LEDs are turned on To avoid confusion it recommended to change settings with the LEDs turned off 4 3 6 Hand Display Screen Orientation In this menu the LCD orientation can be rotated 180 This allows...

Страница 17: ...cted to the internet the driver will install automatically For manual installation the driver is available as a software download from the Excelitas website Note that administrator privileges may be r...

Страница 18: ...ch Windows update for the software Click Next to continue f Select Install from a list or specific location Advanced to locate the driver and click Next Select Include this location in the search and...

Страница 19: ...toolbar 5 1 5 X Cite Control Panel Tips for Use a X Cite Control Panel is available as a download from the Excelitas website This tool provides an interface for controlling the X Cite mini via PC and...

Страница 20: ...mp Not applicable for X Cite mini 5 1 6 Commercial Software Support a X Cite mini can be controlled through many commercially available packages Where X Cite mini specific control may not be available...

Страница 21: ...energy and mechanical components accumulate wear To help minimize unnecessary consumption at the end of an imaging session or if equipment is accidentally left on X Cite mini is equipped with a TTL t...

Страница 22: ...turn the LED on and off The foot pedal is an optional accessory and can be purchased separately from Excelitas Technologies 1 Installing the foot pedal a Locate the FOOT PEDAL IN FP IN input port on...

Страница 23: ...ng temperature Ensure the room temperature is within the recommended operating limits If unit has been stored transported in a cold environment allow it to warm up to room temperature and restart If t...

Страница 24: ...ove the fuse from the tray and check with a multi meter set to resistance An intact fuse will read 0 or another very low value an open fuse will have an extremely high reading d Note Fuses that chroni...

Страница 25: ...roll through the menu options Turn dial in the opposite direction Lock icon is on display speedDIAL not responding to manual control Send unlock command via computer OR power down unit and restart TTL...

Страница 26: ...f X Cite mini s LED Head miniCUBE and speedDIAL can be cleaned with a mild soap and water solution and lint free cloth 1 Turn unit off and disconnect AC power prior to cleaning 2 Use a damp cloth only...

Страница 27: ...ral care and maintenance guidelines described above are followed no further maintenance is required to ensure continued superior performance If through accident or negligence the cable between miniCUB...

Страница 28: ...Storage Conditions Temperature 35 to 60 C Relative Humidity 10 to 95 RH non condensing Atmospheric Pressure 500 to 1060 hPa 8 5 Input Output I O Connections Connection Connection Style Purpose speedDI...

Страница 29: ...tibility The X Cite mini has been tested and found to comply with product safety and electromagnetic compatibility requirements For a complete list of tests and for certification details please contac...

Страница 30: ...generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this...

Страница 31: ...The crossed out wheeled bin symbol indicated above invites you to use those systems If you need more information on the collection reuse and recycling systems please contact your local or regional wa...

Страница 32: ...valid This guarantee may not form the basis for any claims for damages in particular not for compensation of consequential damages The warranty is not transferable No warranty is extended to perishab...

Страница 33: ...8752 USA and Canada Fax 905 821 2055 1 800 668 8752 USA and Canada x cite excelitas com https www excelitas com product category x cite illuminators 11 2 Technical Support and Service For Technical Su...

Отзывы: