27
X-Cite® mini+ User Guide
035-00587R rev. 5
d.
Repeat the previous step with a fresh portion of lens tissue (or new cotton swab) – this
will help avoid recontamination of the optical surface and minimize the amount of
cleaning required.
e.
Let solvent evaporate and verify that the optical surface is clean. Repeat cleaning steps
as necessary.
f.
Before reinstalling and/or using the optics, allow them to dry completely.
CAUTION: Before using any solvent, consult the manufacturer’s Material Safety Data Sheet (MSDS)
and your internal Health and Safety Advisor for proper handling, storage, and disposal instructions.
7.4
Thermal Management System
X-Cite® mini+’s unique thermal management system incorporates air cooling technologies. If the
general care and maintenance guidelines described above are followed, no further maintenance is
required to ensure continued superior performance.
If, through accident or negligence, the cable between miniCUBE and the LED Head is damaged:
1.
Power off and discontinue using the system.
2.
Contact Technical Support (
).
8
Technical Specifications
8.1
General
miniCUBE
LED Head
speedDIAL
Height
110mm (4.4″)
135mm (5.3″)
59mm (2.3″)
Width
180mm (7.1″)
100mm (3.9″)
80mm (3.1″)
Depth
230mm (9.1″)
a
110mm (4.4″)
a,b
112mm (4.4″)
Weight
3.4kg (7.5lbs) miniCUBE and LED Head are
permanently connected
0.3kg (0.7lbs)
Notes:
a.
Does not include clearance required for cable, minimum 200mm (8″).
b.
Does not include microscope flange, varies by type, approximately 10-15mm (0.4-0.6″).
8.2
Electrical
Power Supply
Power Factor Corrected, Universal Input
Input Voltage
100-240VAC, 50/60Hz
Current
XMPS/XMPL: 2.4A max /100V, 1.2A max/240V
XTMS/XTML: 2.4A max /100V, 1.2A max/240V