ACRYLIC CALIBRATION KEY
READ RESULT & RECORD
Result is displayed on the eXact
iDip
®
and in the app . Record
value . For additional tests, repeat
steps 5-9 .
Expect the result to be within 0.15
units from previous calibration .
If variation is greater, please
re-check your ZERO procedure
(steps 5-7) and be sure to use clean water (deionized or
distilled if necessary) .
The Acrylic Calibration Key is a tool for verifying the
calibration of your eXact iDip
®
Photometer compared to
its original factory settings . Run this test immediately upon
receipt and record value for future reference . Follow steps
1-5 on
pages 16-18
. Then, follow the remaining steps below .
DIP KEY AND PRESS READ
Place the Acrylic Calibration
Key into the water sample in the
center of the CELL . Be sure the
key is positioned upright (vertical)
and to the bottom of the CELL .
Press READ .
ZERO METER
Press
and the eXact iDip
®
display reads 0
PPM
indicating the
meter is ready for testing .
SELECT TEST
Tap SELECT TEST at
the top and select the
Calibration Key test .