Examination of the Photometer
The verification standard kit for the P7 Advanced is designed to assure
the user of the accuracy and the reliability of the results related to the
integrated wave lengths.
Order code Verifikation-Standard-Kit: W3T330121
The kit contains one zero standard, 3 different vials for checking 3
different wave lengths.
To check the device with a Verifikation Standard Kit, proceed as follows:
Press the [ZERO/TEST] key and
hold
.
Switch the unit on using the [ON/OFF] key.
Allow the 3 decimal points to be displayed before releasing the [ZERO/
TEST] key.
The instrument displays for a short time
followed by a measurable wavelength.
Scroll through wavelengths using the [MODE] key.
Display
Wavelength
A530
530 nm
A560
560 nm
A610
610 nm
Place the supplied blank ("Zero" Verifikation Standard) in the sample
chamber making sure that the marks are aligned .
Place the supplied Verifikation Kit cover on the adapter.
Press the [ZERO/TEST] key.
The symbol flashes for approx. 8 seconds.
The display shows:
On
Off
Zero
Test
A530
Zero
Test
0.0.0
AbS
. . .
Mode
Calibration Mode
Содержание WALLACE & TIERNAN P7 ADVANCED
Страница 1: ...WALLACE TIERNAN PHOTOMETER PHOTOMÈTRE P7 ADVANCED BETRIEBSANLEITUNG INSTRUCTION MANUAL MODE D EMPLOI ...
Страница 33: ...33 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 63: ...63 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 93: ...93 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 94: ...94 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 95: ...95 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 96: ...96 WT 050 311 000 DM IM 0119 P7 Advanced ...