Operation
Switch the unit on using the [ON/OFF] key.
The display shows the following:
Select the required test using the [MODE] key.
Scroll Memory (SM)
To avoid unnecessary scrolling for the required test method, the instru-
ment memorizes the last method used before being switched off. When
the instrument is switched on again, the scroll list comes up with the
last used test method first.
The display shows the following:
Fill a clean vial with the water sample up to the 10 ml mark, screw
the cap on and place the vial in the sample chamber making sure that
the marks are aligned.
Press the [ZERO/TEST] key (see OTZ).
The "Method" symbol flashes for approx. 8 seconds.
The display shows the following:
After zero calibration is completed, remove the vial from the sample
chamber. The characteristic coloration appears after the addition of
the reagents.
Replace the cap on the vial and place in the sample chamber making
sure that the marks are aligned.
Press the [ZERO/TEST] key.
(For Countdown/reaction period see page 37)
The "Method" symbol flashes for approx. 3 seconds.
The result appears in the display.
The result is saved automatically.
Repeating the test:
Press the [ZERO/TEST] key again.
OTZ (One Time Zero):
The zero setting is held in memory until the instrument is switched off.
It is not necessary to perform a new zero each time, if the water sam-
ples under test are from the same body of water and the conditions
of testing are the same.
The zero setting can be repeated each time if necessary.
Repeating the zero:
Press the
[ZERO/TEST]
key for 2 seconds.
METHOD
METHOD
METHOD
RESULT
On
Off
Mode
Zero
Test
Zero
Test
Zero
Test
0.0.0
Zero
Test
METHOD
Содержание WALLACE & TIERNAN P7 ADVANCED
Страница 1: ...WALLACE TIERNAN PHOTOMETER PHOTOMÈTRE P7 ADVANCED BETRIEBSANLEITUNG INSTRUCTION MANUAL MODE D EMPLOI ...
Страница 33: ...33 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 63: ...63 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 93: ...93 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 94: ...94 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 95: ...95 WT 050 311 000 DM IM 0119 P7 Advanced ...
Страница 96: ...96 WT 050 311 000 DM IM 0119 P7 Advanced ...