installation, and equipment for industrial use and some other equipment, for example, stationary
motors with permanent connection to the fixed installation.
CAT IV: For measurements performed at the source of low voltage installation, i.e. electricity meters.
1.0 Introduction / Scope of Supply
The instruction manual contains information and references, necessary for safe operation and
maintenance of the instrument. Prior to using the instrument (commissioning/assembly) the user is
kindly requested to thoroughly read the instruction manual and comply with it in all sections. Failure to
read the instruction manual or to follow with the warnings and references contained herein can result
in serious bodily injury or instrument damage.
EVOMEX Solarmex 1000 mem is constructed for the easy to use test and measurements of
photovoltaic modules. The terms used in these operating instructions to destinate systems for
generating electrical power from sunlight include: solar system, photovoltaic system, PV system and
the like.
• Voltage Measurement up to 1000V DC
• Short Circuit Measurement up to 20A DC
• Insulation Measurement Voltage 250V/500V/1000V DC
• Insulation Measurement up to 20 M
Ω
with limit indication
• Continuity Test of Protective Earth
• Earth Fault Measurement
• Polarity Test
• LC Display with Back Light
• Optional Temperature Measurement
Scope of Supply:
1x EVOMEX solarmex 1000
4x Batteries 1,5V IEC LR6, AA
4x Safety Test Leads 1.5m red, blue, yellow
1x Safety Test Probe
1x Safety Crocodile Clamp
2x PV Adaptor red
2x PV Adaptor blue
1x Carrying Case with foam inlay
1x USB Interface cable USB A - USB Mini B
1x Instruction Manual
2.0 Transport and Storage
Please keep the original packaging for later transport, e.g. for calibration. Any transport damage due to
faulty packaging will be excluded from warranty claims.
Instruments must be stored in dry and closed areas.
In the case of an instrument being transported in extreme temperatures, a recovery time of
minimum 2 hours is required prior to instrument operation.
3.0 Safety
The instruments have been constructed and verified in compliance with the latest safety standards for
Test Instruments IEC/EN 61010-1 and IEC/EN 61557 and have left the factory in safe and perfect
conditions.
In order to avoid electrical shock, the valid safety and VDE regulations regarding excessive
contact voltages must receive utmost attention, when working with voltages exceeding 120V (60V) DC
or 50V (25V)rms AC. The values in brackets are valid for limited ranges (as for example medicine and
agriculture).
Prior to usage ensure perfect instrument function.