IE-SNC-Tx5xx-03
8
Checking operation
Visual check
After connecting the power supply the LCD displays a currently measured value (if the LCD
has not been switched off). If the measured value exceeds the measuring range of the temperature
probe or the temperature probe is not connected correctly Err 1 or Err 2 reading appears on the
LCD. In this case check the temperature probe connection and that the value being measured is in
the allowed measuring range. (Connection of the temperature probe is relevant only for the T4511
transducer).
Communication check via a LAN
Open a browser of www pages and enter the device IP address.
Example
:
http://192.168.1.213
The device displays its name, serial number, measured values, alarm states and alarm
settings (Figure 5). If „Access denied“ appears instead, the display of www pages is not allowed
(Figure 10). (Change the appropriate setting in the Setup - more details in chapter WWW- WWW
configuration, page 20).
Figure 5 – device www pages
Calibration
Recommended calibration interval is
•
one year for T3511