E2608-VOC
Rev 05.03.2021
21
0x0005
5 / 40006 RW Baud rate *
1200, 2400, 4800, 9600, 19200,
38400, 57600
9600
0x0006
6 / 40007 RW Response delay, ms
1...255
10
0x0007
7 / 40008 RW Stop bits, parity bit *
1 – No parity bit, 1 stop bit
(default after factory reset)
2 – No parity bit, 2 stop bits
3 – Odd parity, 1 stop bit
4 – Even parity, 1 stop bit
NOTE: 3 and 4 are available
starting from the Software
version 0x218 (dec. 536)
1
0x0008
8 / 40009 R Last error code
1...255
-
0x0011 17 / 40018 RW
Technological:
age of last data in seconds
(read) / restart (write)
0...65535 s (read),
writing 42330 restarts the
instrument
-
Address
Reg / MHR RW Description
Supported values (dec)
Default
0x00A5 165 / 40166 RW Zero adjustment for gas
data, ADC
-32000...+32000 ADC units
0
0x00A6 166 / 40167 RW Slope adjustment for
gas data
1...65535
512
0x00A7 167 / 40168 RW Change rate limit for gas
data, units / s
1...32000, 0 - no limit
0
0x00A8 168 / 40169 RW Integrating filter time
constant, s
1...32000 (seconds),
0 - no filter
0
0x00C9 201 / 40202 RW
Parameter tied to analog
output 1
0 – None
2 – Gas concentration
9 – Forced Modbus control,
value set in MHR / 40204
2
0x00CA 202 / 40203 RW
Parameter tied to analog
output 2
0 – None
2 – Gas concentration
9 – Forced Modbus control,
value set in MHR / 40205
2
0x00CB 203 / 40204 RW Forced value for analog
output 1***
0...1000
(0.0%...100.0% of output scale)
0
0x00CC 204 / 40205 RW
Forced value for analog
output 2***
0...1000
(0.0%...100.0% of output scale)
0
Содержание PluraSens E2608-VOC
Страница 1: ...PluraSens Solvent Vapor Detector Transmitter E2608 VOC User Manual ...
Страница 3: ...E2608 VOC Rev 05 03 2021 Modbus holding registers 20 Warranty 25 Manufacturer contacts 25 3 ...
Страница 14: ...E2608 VOC Rev 05 03 2021 Mounting dimensions Wall mount version with mounting lugs Duct mount version 14 ...