E2658-CO

Rev 06.04.2021

Operation

Turn on the power. The instrument warm-up time takes about 1 minute after switching on
and the final sensor stabilization time to maximum accuracy takes about 30 minutes. The
operating status is indicated by the LED on the PCB of the device. The control LED (red)
response to different processes is presented in the following table:

Mode

LED mode

During calibration mode or sensor heating
period (if activated)

0.5 Hz (50%  on, 50% off)

Relay 1 turned on

Blinking 1 Hz (50% on, 50% off)

Relay 2 turned on

Blinking 2 Hz (50% on, 50% off)

During the Modbus communication cycle

Short on-off pulses

Normal operating/measurement

Continuously on or off

Make sure that the detector-transmitter is properly mounted, the external devices
connected, the power LED (green) on, and the control LED (red) is constantly lit. Place the
lid back and fix it with the screws. Make certain that the cable gland and the screws fixing
the front lid are properly tightened to ensure the conformity to IP66 protection class and
ATEX requirements. Make certain that the device is properly earthed. The device is ready to
use.

Maintenance

Do not perform any maintenance operation with the power on.

Clean the device with a soft damp cloth. Do not use any abrasive cleaning agents. Do not
immerse the device in water or any cleaning media.

Calibration

E2658-CO detector-transmitters have been calibrated by the Manufacturer with standard
gas mixtures before delivery. Provided that the sensor is used under moderate conditions,
field recalibration is recommended every 12 months. Please contact your dealer for more
information.

13

Содержание E2658-CO

Страница 1: ...PluraSens Carbon Monoxide Detector Transmitter E2658 CO User Manual...

Страница 2: ...ns 9 Sensor probe handling 10 Gas sensor replacement procedures 10 Electrical connections 11 Operation 13 Maintenance 13 Calibration 13 Delivery set 14 Order code for E2658 CO options 14 Configuring 1...

Страница 3: ...sive limit UEL vol in air 74 Odor Odorless Hazards Highly toxic Mild poisoning causes lightheadedness confusion headache dizziness and flu like effects Larger exposures can lead to toxicity of the CNS...

Страница 4: ...1 2 gas OUT2 2 gas Output scale width 10 resolution Recommended 20 100 of the detection range Load resistance RL Us 2 V 22 mA for 4 20 mA RL 250 kOhm for 0 10 V mode Digital interface RS485 Modbus RTU...

Страница 5: ...9 1 2016 EN 60079 29 2 2015 EN 60079 29 3 2014 Operating conditions 20 50 C 0 90 RH non condensing 0 9 1 1 atm Explosion safe zones and ATEX Zones 2 and 22 Non aggressive atmosphere NOTE The sensor ex...

Страница 6: ...r power supply The version of your detector is marked on the package If the symbol is marked on the equipment consult the documentation for further information Safety requirements Misuse will impair t...

Страница 7: ...the installation During the installation of the detector transmitter the following points must be considered Application air quality control or leakage detection Properties of the space under investi...

Страница 8: ...he sealed area of the device see dimensional drawing Unscrew four lid screws and detach the lid from the detector Fix the detector on the wall and earth the enclosure using an earthing terminal on the...

Страница 9: ...E2658 CO Rev 06 04 2021 Mounting dimensions Front view Bottom view 9...

Страница 10: ...since this may damage the sensor Do not remove the filter as it may cause the device to show incorrect values and or break the sensor The recommended orientation of the sensor probe is vertical with t...

Страница 11: ...r supply and of the external devices Without turning on the power plug the power cable and connect the analog relay outputs and or digital interface terminals to the necessary devices according to the...

Страница 12: ...to an external RS485 network Both analog outputs can be independently changed between 4 20 mA and 0 10 V type using jumpers J1 OUT1 and J2 OUT2 By closing pins on a specific jumper the related output...

Страница 13: ...s properly mounted the external devices connected the power LED green on and the control LED red is constantly lit Place the lid back and fix it with the screws Make certain that the cable gland and t...

Страница 14: ...r E2658 Mounting accessories 2 screws with plastic dowels Order code for E2658 CO options E2658 options Order code Integrated 90 265 V mains power supply module E2658 CO 230 Integrated 24 VAC power su...

Страница 15: ...xible setting of analog output scales for each output Output zero and slope adjustment for calibration Free assignment of each of two relays to the chosen parameter Several relay control logic modes S...

Страница 16: ...ter Error codes 01 Illegal function 02 Illegal data address 03 Illegal data value 04 Slave device failure details of last error 04 can be read from register 0x0008 Modbus holding registers Register ad...

Страница 17: ...000 32000 ADC units 0 0x00A6 166 40167 RW Slope adjustment for gas data 1 65535 512 0x00A7 167 40168 RW The change rate limit for gas units s 1 32000 0 no limit 0 0x00A8 168 40169 RW Integrating filte...

Страница 18: ...RE1 0 1000 s 0 0x00D8 216 40217 RW Switching delay for relay RE2 0 1000 s 0 0x00D9 217 40218 RW Minimal on off time for relay RE1 0 1000 s 0 0x00DA 218 40219 RW Minimal on off time for relay RE2 0 100...

Страница 19: ...bit 1 0 1 analog outputs deactivated activated bit 2 0 1 in case the sensor is absent turn signaling off on analog output1 bit 3 0 1 in case the sensor is absent turn on signaling with low current hig...

Страница 20: ...00 value for analog output 1 32000 32000 gas units User defined 0x0107 263 40264 RW 0 value for analog output 2 32000 32000 gas units User defined 0x0108 264 40265 RW 100 value for analog output 2 320...

Страница 21: ...anufacturer will at its option either repair or replace a product that proves to be defective This warranty is void if the product has been operated in conditions outside ranges specified by the Manuf...