Alarm
Reason for alarm
What should I do?
Occlusion
The infusion route (reservoir,
infusion set, cannula) is blocked.
Press
to acknowledge the alarm and
remove the infusion set from body. Connect
a new infusion set to the pump, as described
in
Section
4.3
, and use the priming
function. Treatment can be continued once
the drug is visible in the tube or comes out
the end of the tube during the priming cycle.
If the tube becomes blocked again during
priming or at a later point, dispose of the
reservoir with the household waste as
described in
Section
9.5
and start filling a
new reservoir. Proceed here in accordance
with the instructions in
Section
4.4
.
Alarm
Reason for alarm
What should I do?
Device error
The red signal light
to the right above
the screen will
flash.
The internal pump monitoring system has
detected a technical error (e.g., sensor
failure, etc.). Drug delivery has stopped.
1. Replace the rechargeable battery
An empty or defective battery is the
most likely root cause of a device error.
Therefore, replace the rechargeable battery
first and check to ensure that the new
battery is adequately charged and properly
connected.
2. Contact Customer Service
If the device error is unable to be resolved
by replacing the rechargeable battery, then
contact Customer Service.