YOUR
EVER PHARMA D-
mine
®
PUMP IN EVERYDAY LIFE
Description
What should I do?
The infusion set
has been primed
correctly, delivery
has started, the
drug effect fails
to appear, the
device signals an
"OFF" state
First, check whether the threaded connection between the pump and infusion set is tightened firmly and
the infusion set is properly connected to the body.
If this is the case, but the problem persists, stop drug delivery and disconnect the infusion set from body.
Remove the reservoir and the infusion set and dispose of both. Remove the rechargeable battery from the
device.
Inspect the pump and rechargeable battery for cracks or damage. Restart the device and monitor the
start‑up process. Each step must be properly displayed on the screen.
If there is no visible damage on the pump and rechargeable battery, then you can continue treatment with a
new reservoir. Change the reservoir and infusion set in accordance with the instructions in
Section
3.1
.
If the pump is defective, contact Customer Service.
Incorrect screen
display
If the buttons cannot be unlocked, then briefly remove the rechargeable battery from the pump and
reinsert.
If the pump is defective, contact Customer Service.
Air in the
reservoir
visible after filling
Check the filling level of the reservoir in the viewing window after the filling process. Air bubbles can be
removed from the reservoir by a priming process via the connected infusion set. To do this, follow the
instructions for priming in
Section
3.2
No liquid in the
infusion set
Check the filling level of the reservoir before connecting the infusion set with the body. If no liquid reaches
the infusion set even after priming process, repeat the process with a new reservoir and new vial.
Instructions in
Section
3.1