Figure 56 : DO (mg/L) – Page 1
This page allows you to set operating parameters for the DO Concentration
measurement mode:
Parameter
Function
Factory
Default
Select DO Unit
Sets the unit of measurement for DO
Available units: mg/L and ppm
mg/L
Auto Salinity Comp
DISABLE
– No automatic compensate
ENABLE
– The meter will apply measured salinity for DO
concentration of your sample.
Note:
This option is activated only in CD650 & PCD650.
DISABLE
Set Salinity
Allows you to set the salinity (in ppt) for your solution
Available range: 0 to 50 ppt
0.0 ppt
DO (mg/L) – Page 2
NOTE:
Auto salinity compensation will be applicable only in multi measurement
mode and only if one of the measurement selected is conductivity related. Otherwise,
a manually set salinity value will be applied.