105
Function Keys available in saturation calibration mode:
TEMP (F1)
Goes to temperature calibration
CLR-C (F2)
Clears previous calibration (if any) after
ENTER
key is pressed in
confirmation screen
NEXT (F3)
Shows calibration report
ESC (F4)
Exits from calibration and goes back to saturation measurement
mode
ENTER
Confirms the calibration
(Not functional)
2.3 To calibrate 0% saturation
Figure 53 : 0% Saturation Calibration Screen
1. Press
MODE
key to select % Saturation mode.
2. Immerse the DO probe in 0% solution. Stir gently to create a
homogenous solution.
3. Press
CAL
key to calibrate the meter. The meter shows the ‘Dissolved
O
2
Calibration-Rinse Electrode’ screen for a few seconds to prompt
user to rinse electrode with de-ionized water before calibration.
4. The meter then shows saturation calibration screen. Wait for the
reading to stabilize.
5. The primary display will show the current value of measurement.
6. Press
ENTER
key to confirm the calibration. The meter automatically
calibrates to 0% saturation and returns to the measurement mode.
Note:
If you wish to completely re-calibrate the meter, you need to clear
previous calibration data using the
CLR-C (F2)
key. The meter shows the
confirmation screen. Press
ENTER
key to confirm deleting of the previous
calibration. It will
clear
all the calibration data including mg/L data.
The keys that appear in 0% saturation calibration mode function similarly to
the one as described in 100% saturation mode.
Содержание CD 650
Страница 2: ......
Страница 4: ......
Страница 10: ...2 ...
Страница 28: ...20 ...
Страница 45: ...37 ...
Страница 49: ...41 ...
Страница 51: ...43 ...
Страница 61: ...53 ...
Страница 65: ...57 ...
Страница 67: ...59 ...
Страница 73: ...65 ...
Страница 82: ...74 END OF CONDUCTIVITY SECTION ...
Страница 84: ...76 ...
Страница 92: ...84 ...
Страница 98: ...90 ...
Страница 100: ...92 ...
Страница 106: ...98 ...
Страница 108: ...100 ...
Страница 126: ...118 ...