15
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD
5.4 HOLD
Current
Setting
When Transmitter is in CAL or SETUP modes, it automatically goes into a
‘HOLD’ mode.
To indicate Transmitter is in ‘HOLD’ mode, output current can be set to
22
mA
output by activating the ‘HLD On’.
1. Press ENT key and use
▲
or
▼
keys to scroll till LCD displays
‘
SEt
’ in the upper display; and ‘
HLd
’ in the lower display. Press
ENT again.
2.
Upper display now shows ‘
HLd
’. Lower display will show either
‘
OFF
’ or ‘
On
’. Use
▲
or
▼
keys to toggle between ‘
On
’ of ‘
OFF
’.
Press ENT to accept selection.
3. Press
S
and
T
keys together, to return to MEAS mode.
NOTE: If ‘HLd’ is set to ‘OFF’, then current output will be equivalent to last
measured value.
5.5 Out-of-Range
Current
Setting
If pH or ORP readings exceed the scale, LCD displays ‘Or’ (over-range), or
‘Ur’ (under-range). Transmitter current output can be set to
3.8 mA
to indicate
Out-of Range situation.
1. Press ENT key and use
▲
or
▼
keys to scroll till LCD displays
‘
SEt
’ in the upper display; and ‘
org
’ in the lower display. Press
ENT again.
2.
Upper display now shows ‘
org
’. Lower display will show either ‘
OFF
’
or ‘
On
’. Use
▲
or
▼
keys to toggle between ‘
On
’ of ‘
OFF
’. Press
ENT to accept selection.
3. Press
S
and
T
keys together, to return to MEAS mode.
NOTE: If ‘org’ is set to ‘OFF’, under-range will be set at 4 mA, while over-
range will be set at 20 mA.
SETUP
HOLD
SETUP
HOLD
SETUP
HOLD