9
4.2 pH
Calibration
A complete 2-point CAL is required for this Transmitter. If Calibration is
aborted after 1
st
calibration point, transmitter reverts to previous
Calibration data.
1. Rinse probe thoroughly with de-ionized water or a
rinse solution. Blot excess liquid.
2. Dip probe into calibration buffer. End of probe must be
completely immersed into buffer. Stir probe gently to
create a homogeneous sample.
3. From pH MEAS mode, press CAL to enter pH
calibration mode. CAL indicator will be shown. Primary
display will show measured reading while secondary
display will indicate the pH standard buffer solution.
4. Press
▲
or
▼
keys to select either pH 7.00 (USA) or
pH 6.86 (NIST) standard buffers. This is the first point,
Offset calibration.
5. Wait for measured pH value to stabilize. Press
ENTER to confirm calibration. Transmitter is now
calibrated to current buffer.
6. Rinse probe with de-ionized water or rinse solution,
and place in next pH buffer (pH 4.01, pH 9.18 or pH
10.01).
7. Repeat steps 5 and 6.
8. Upon successful Calibration, transmitter displays
Slope (primary display, in mV) and Offset (secondary
display, in pH) of electrode.
9. Press ENTER to return to MEAS mode. Transmitter
has been calibrated. Rinse electrode and proceed
with measurements.
NOTES:
To exit from pH calibration mode without confirming calibration, press
▲
and
▼
keys
together.
If selected buffer value is not within
±
1.0 pH from measured pH value: electrode and buffer icon blink
and ERR annunciator appears in lower left corner of display.
CAL
HOLD
pH
READY
.
.
pH
CAL
HOLD
pH
READY
.
.
pH
CAL
HOLD
pH
READY
.
.
pH
CAL
HOLD
pH
.
.
mV