- 22 -
P3.3: Select temperature compensation
This controller features selectable automatic temperature compensation (ATC) or
manual temperature compensation (MTC). ATC automatically compensates for
temperature fluctuations; MTC let you select a specific value at which temperature
will be compensated. ATC readings require a temperature sensor.
1.
Key in the password “022” as per procedure in Section 5.
2.
Press
▲
key twice. The screen will scroll P3.0 and CONF. Press ENTER
three times.
3.
The screen will scroll P3.3 and MTC (or ATC).
4.
Press
▲
or
▼
keys to toggle between MTC and ATC.
5.
Press ENTER to confirm and return to general Setup mode.
Press
▲
and
▼
keys together (ESCAPE) to return to measurement mode.
NOTE: ORP is not affected by temperature and so ATC does not operate in ORP
mode. When set for ATC, the pH 200 will display temperature if a 100
Ω
Pt 100
sensor is connected.