After diagnostic routine, the calibrator will be forced into the "IN" function with the active parameter
previously selected (i.e. with the indication of fig.8.5.1.B).
I N 0 . 0 0 mA
Fig. 8.5.1B
Press <
IN/OUT
> key to select the simulation mode with an indication as per fig. 8.5.1.C.
OUT 0 . 0 0 mA
Fig. 8.5.1C
Press the <
s
> or <
t
> keys to adjust the simulation value.
A single touch on one of the two
<
s
> and <
t
> keys will adjust the simulated value of one least significative digit.
A continuous touch will cause repetitive increments of the simulated value.
The speed of variation can be accelerated by pressing <
FAST
> + <
s
> or <
FAST
> + <
t
> keys; a
single touch will be equivalent to the third significative digit (L.S.D.A. x 100)
8.5.2
Parameter selection
To select the electrical parameter in measuring or simulation mode, follow the procedure indicated
below:
♦
Switch the instrument -ON-.
♦
Press the <
SELECT
> key; the display will show one of
the menu pages, indicated in figure.
♦
Press <
s
> or <
t
> key to select the appropriate page.
♦
Press the <
SELECT
> key to invalidate the new
instruction or the <
ENTER
> key to load the selection in
the memory; the instrument will return to the previous
function with the new selected mode (IN or OUT).
8.5.3
Average readings
The use of the "Average" function is advised with unstable input signals.
The average represents a progressive integration of the input signal on the last 32 conversions
(approximately 10 seconds).
Select the measuring mode (-IN-)
Se l : 4 - 2 0 mA
Se l : 4 - 2 0 Ñ mA
Se l : 4 - 2 0 S mA
Se l : 0 - 2 0 S mA
Se l : 0 - 2 0 EX mA