_______________________________________
______________________________________
Instruction Manual MM850099 ed.02a
_________________________________________ 21 ________________________________________
ON
To power the instrument actuate the <
ON
> key (fig. 8.3.A); the indication of fig.
8.3.B will appear for a few seconds.
™. . . Un i Ca l . mA. . . ™
Fig. 8.3B
With power -ON- , the instrument will run an autodiagnostic routine for the self-checking of critical
circuits and components.
A positive check will be shown with the indication of fig. 8.3.C for about one second,
Te s t OK Ve r 2 . 0 0 0
Fig. 8.3C
The number of the right site of the display indicates the version of the program installed on the
instrument.
I N 0 . 0 0 mA
Fig. 8.3C
The instrument will go to the previously selected operating mode, i.e. as per fig. 8.3.D.
Faulty conditions will be indicated as described in par. 8.8.
8.4
Battery voltage indication.
During normal operating modes (measure or simulation) "low battery" condition will be shown as per fig.
8.4.A
I n 0 . 0 0 mA
Fig. 8.4A
An empty symbol means that the battery has enough energy for about 30 minutes of operation.
A black symbol means that the battery charge level is below the minimum: batteries must be recharged
for a full period of 12 hours.
8.5
Operating mode set-up
To select the required operating mode follow the procedures indicated below.
8.5.1
IN / OUT function selection
ON
Switch the instrument -ON