![Eurotherm Chessell 4100G Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/eurotherm-chessell/4100g/4100g_technical-manual_2440593031.webp)
100mm GRAPHICS RECORDER TECHNICAL MANUAL
HA249875
Issue 1 Jan 99
Section 2
Page 2 - 10
2.3.3 RESISTANCE INPUTS (Cont.)
Figure 2.3.3 Calibration check wiring: resistance inputs
Table 2.3.3 Resistance input calibration checkpoints.
0 to 150
Ω
0 to 600
Ω
0 to 6k
Ω
Input range (
Ω
)
Check points (
Ω
)
Max tol.
(
Ω
)
0
17
33
50
67
84
100
117
133
150
0.1145
0
67
133
200
267
333
400
467
533
600
0.437
0
667
1333 2000
2667
3333
4000 4667 5333
6000
4.145
1
2
3
4
5
6
7
8
9
10
V+
V-
I
To further
channels
Resistance
box
V+
V-
I
V+
V-
I
V+
V-
I
V+
V-
I
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание 4100G
Страница 3: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 53: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 55: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 57: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Страница 61: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...