6-36
6.9
Test Signal Level
0.1 Vrms±10%
0.3 Vrms±10%
1.0 Vrms±10%
6.10
Output Impedance
30
Ω
±5%
100
Ω
±5%
6.11
Display Range
Parameter
Frequency
Measuring
100Hz
、
120Hz
1
μ
H to 99999H
1kHz
0.1
μ
H to 99999H
L
10kHz
0.01
μ
H to 99999H
100Hz
、
120Hz
1pF to 99999
μ
F
1kHz
0.1pF to 99999
μ
F
C
10kHz
0.01pF to 99999
μ
F
R,Z
0.1m
Ω
to 99.9M
Ω
Q
0.0001 to 99999
D
0.0001 to 9.9999
6.12
Correction Function
Open Correction
Open correction eliminates measurement errors due to parasitic stray impedances of the
test fixture.
Short Correction
Short correction eliminates measurement errors due to parasitic residual impedances of the
test fixture.
6.13
Comparator Function
U2810D’s built-in comparator can short devices into a maximum of 4 bins (P1, P2, P3 and
NG) using a maximum of three pairs of primary limits and one pair of secondary parameter
limits. Also, a device whose primary parameter is within limits, but whose secondary
parameter measurement result is not within limits can be sorted into an AUXiliary bin. The
comparator function is especially useful when using the U2810D with a component handler.