3-14
displayed on the LCD.
3. Repeat step 2 and 3, signal source output impedance will be switched between 30
Ω
and 100
Ω
.
3.2.7 Measurement Range
Perform the following steps to set the measurement range
1. Assume the current measurement range is set to “Auto” status. Press
RANGE
key to
change the measurement range from the “Auto” mode to the “Hold” mode. When the
measurement range is set to the “Hold” mode, “
Auto
” is turned off from the LCD, the
impedance range is fixed at the current range setting.
3.2.8 Trig Mode
U2810D provides 4 kinds of trig modes:
INT : Internal trigger mode
MAN: Manual trigger mode
EXT : External trigger mode
DUT : DUT trigger mode
BUS: BUS trigger mode
Performing the following steps to set the measurement mode
:
1. Press TRIG MODE key to switch trig mode from INT to BUS, and the current
measurement mode will be displayed on LCD.
Note: when the DUT mode is selected, the nominal value of primary parameter should be
entered.
3.2.9 Correction
U2810D’s OPEN correction capability cancels errors due to the stray admittance (G, B) in
parallel with the device under. U2810D’s SHORT correction capability corrects for the
residual impedance (R, X) in serial with the device under test.
Perform following steps for the open and short correction:
1. When the U2810D is under the measurement status, Press CLEAR key to enter the
correction function.
2. If the fixture is open, the information shown in Figure 3-2 will be displayed.