48
© Emerging Tech Trans, LLC
7.
Using tweezers take the small “clear”
valve/gasket from each valve and store them in a
separate container
–
There should be 2 per stimulus channel. Depending on the odorant
carrier version there may be one or two additional valves in the flush channel
8.
Place the whole carrier top into the cleaning bath (WARNING: This may remove the top
sticker on the odorant carrier). An alternative to full submersion is to thoroughly wipe the
underside and gasket receptacles with alcohol soaked, lint free cloth.
Содержание Olfactometer 2
Страница 1: ...ETT Olfactometer ETTDirectControl The First Paradigm Manual and Tutorial Revision 2021 1 ...
Страница 2: ...2 Emerging Tech Trans LLC ...
Страница 52: ...52 Emerging Tech Trans LLC Odorant flow setup Carrier Setup Check valve layout in the odorant carrier ...
Страница 53: ...53 Emerging Tech Trans LLC Typical air flow configuations ...
Страница 54: ...54 Emerging Tech Trans LLC ...
Страница 55: ...55 Emerging Tech Trans LLC Configuration for multiple channels at once ...
Страница 56: ...56 Emerging Tech Trans LLC Setting up multiple channels within ETTDirectControl ...