© Emerging Tech Trans, LLC
Cleaning the odorant carrier
Before we start:
1.
The odorant carrier is made from soft solid Teflon material
i.
Only hand tighten screws
ii.
If screws are too tight, they will strip the Teflon block
iii.
If screws are not tight enough, they will not seal
2.
The odorant carrier is sealed with small clear silicone flap valves/gaskets
i.
They can be hard to see on light colored surfaces
ii.
Small debris can prevent them from sealing properly when reinstalling
iii.
Do not lose
3.
A cleaning bath will be required
i.
A cleaning bath will be required
ii.
The choice of detergent or solvent depends on the odorants used
iii.
Alcohol or non-odorous soap are typical
For questions and advice consult emergingtechtrans.com in the support section or contact ETT at
Cleaning the carrier top
This procedure is only recommended for experienced users and comes at the risk of damaging/removing
odorant carrier labels and stickers. An alternative procedure is to only soak valves and gaskets as
described in the below process but not submerge the entire carrier.
1.
Unscrew the bottles from the carrier top
2.
Cover the bottles with their respective cap
3.
Put a clean pad or towel on the clean work surface and put the carrier top on it
Содержание Olfactometer 2
Страница 1: ...ETT Olfactometer ETTDirectControl The First Paradigm Manual and Tutorial Revision 2021 1 ...
Страница 2: ...2 Emerging Tech Trans LLC ...
Страница 52: ...52 Emerging Tech Trans LLC Odorant flow setup Carrier Setup Check valve layout in the odorant carrier ...
Страница 53: ...53 Emerging Tech Trans LLC Typical air flow configuations ...
Страница 54: ...54 Emerging Tech Trans LLC ...
Страница 55: ...55 Emerging Tech Trans LLC Configuration for multiple channels at once ...
Страница 56: ...56 Emerging Tech Trans LLC Setting up multiple channels within ETTDirectControl ...